Page: 61 of 102
Status
Int.Err
Desicc.
Humid
Reading
0..49/EE
System-Alarm-LED
Red/Steady
Green/ Blinking
Alarm Circuit 1,2
Alarm
Normal Operation
Reason
Internal error.
Measuring
Cell
Humidity critically
high (>40%rH)
Sample
temperature below
dew point.
Instantaneous
condensation
possible.
Recommendations
Restart the system.
Check/
Replace
desiccator.
Check/
Replace
desiccator.
Status
Com?
Sample?
Data log?
Reading
EE
0..49/EE
System-Alarm-LED
Red/Steady
Alarm Circuit 1,2
Alarm
Reason
No communication
between Computer
Unit and Measuring
Cell.
Meter is not able to
measure the
sample:
no water in, oil
content much too
high no light
transmission
possible.
Data logging is not
possible:
No DECKMA
memory card
inserted.
A read only card has
been inserted.
-A new DECKMA
memory card has
been inserted but
has not been
activated.
Recommendations
Check
connection
between Computer
Unit and Measuring
Cell.
-Check sample,
check sample flow
rate, clean Sample
Glass Tube
according to section
12.
Insert the active
memory card.
New memory card.
Activate the
memory card.
Содержание OMD-2008
Страница 28: ...Page 27 of 102 5 5 Cabling OMD 2008 EV Figure 9 Bottom View OMD 2008 EV ...
Страница 84: ...Page 83 of 102 ...
Страница 85: ...Page 84 of 102 ...
Страница 86: ...Page 85 of 102 ...
Страница 87: ...Page 86 of 102 ...
Страница 88: ...Page 87 of 102 ...
Страница 89: ...Page 88 of 102 ...
Страница 90: ...Page 89 of 102 ...
Страница 91: ...Page 90 of 102 ...
Страница 92: ...Page 91 of 102 ...
Страница 93: ...Page 92 of 102 ...
Страница 94: ...Page 93 of 102 ...
Страница 95: ...Page 94 of 102 ...
Страница 99: ...Page 98 of 102 Type Approved by ...