16.1 Automatic Cell Cleaning Unit Fault Finding
Fault
Explanation
Solution
Pneumatic
Cleaner not
operating
Air pressure low.
Restore air pressure.
Air connections missing
or wrong.
Sort out air connections. Air
connections are Push-In type. Press
release ring to release hose.
Auto Clean interval set to
“OFF”.
Activate cleaner via menu.
Auto Clean interval set to
long interval, time not
yet expired.
Wait for time to expire, or manually
trigger cleaning cycle. Refer to section
“Change of cleaning frequency”.
Wiper moving
very slowly
Air flow adjuster too
stringent.
Turn Adjuster on Auto Clean Control
counter-clockwise or un-clockwise with
a screw driver.
Wiper or Cylinder not
clean.
Clean wiper.
Wiper moving
very fast
Air flow adjuster too
wide.
Turn Adjuster on Auto Clean Control
clockwise with a screw driver.
Cleaning
insufficient
Cleaner not operating.
Check wiper and air supply.
Debris deposits.
Try additional manual cleaning.
Deposition of material
that cannot be removed
mechanically (e.g. Iron
Oxide.)
Try chemical removal of deposit. Apply
Cleaning Kit. Iron Oxide can be
removed with Citric Acid.
Cleaner frequency not
adjusted to situation on
site.
Adjust cleaner frequency.
Wiper seal worn out or
damaged.
Replace wiper seal.
Wiper piston missing.
Re-install wiper piston.
Please do not hesitate to contact DECKMA Hamburg for additional support.
Содержание OMD-2008
Страница 28: ...Page 27 of 102 5 5 Cabling OMD 2008 EV Figure 9 Bottom View OMD 2008 EV ...
Страница 84: ...Page 83 of 102 ...
Страница 85: ...Page 84 of 102 ...
Страница 86: ...Page 85 of 102 ...
Страница 87: ...Page 86 of 102 ...
Страница 88: ...Page 87 of 102 ...
Страница 89: ...Page 88 of 102 ...
Страница 90: ...Page 89 of 102 ...
Страница 91: ...Page 90 of 102 ...
Страница 92: ...Page 91 of 102 ...
Страница 93: ...Page 92 of 102 ...
Страница 94: ...Page 93 of 102 ...
Страница 95: ...Page 94 of 102 ...
Страница 99: ...Page 98 of 102 Type Approved by ...