- 12 -
HANDPIECE MANUAL MODE
In some situations the Polylase LP laser handpiece detection system can be turned off.
This is useful when handpiece-related errors happen in the system and operator needs to
complete the treatment session. In order to start the laser in handpiece Manual Mode, on
the laser controller Start Screen press and hold “Handpiece Mode” button for 3 sec. The
button status will change from “
Auto
” to “
Manual
” and now the laser can be started with
all handpiece-related errors ignored by the system.
WARNING!
When laser is started in Manual Mode, handpiece spot size can not be
changed as usual or automatically recognized by the laser controller! It is only
possible to work with one currently installed handpiece spot size. If decision was
made to start the laser in Manual Mode, the operator is responsible for the correct
selection of the handpiece spot size (on the laser controller screen) that will match
the actual handpiece output beam spot size. Failure to do so may result in severe
skin burns due to incorrect Fluence settings or other injuries to the patient!
Do not use the laser in handpiece Manual Mode without need and arrange for laser
handpiece service as soon as possible, to eliminate handpiece malfunction that
results in handpiece errors.
SYSTEM SOUNDS
The Polylase LP is equipped with the system sounds function. This includes Pulse Sound
and Error Sound options. The laser starts with both sound options turned OFF by default
but it could be turned ON from the laser
Service
screen.
In order to turn system sounds on, select “Service” button from the
Edit
mode, then turn
the desired sound option (either both or only one) ON by the corresponding buttons on
the
Service
screen.
Содержание POLYLASE LP
Страница 6: ... 2 Figure 1 1 POLYLASE LP Dual Output Laser ...
Страница 26: ... 8 Figure 3 5 Front and side of the POLYLASE LP Laser ...
Страница 32: ... 14 Figure 3 9 Labels rear of the POLYLASE LP Laser Figure 3 10 Handpiece Label Laser Aperture Label 4000 001A ...
Страница 49: ... 10 Figure 5 6 Handpiece protective window is located inside the snap in holder between distance gauges ...