- 11 -
Cleaning the Laser Console
Clean the external surfaces of the laser console periodically with a clean cloth wetted
with a spray cleaner (do not use acetone to clean any Laser Console parts!). Dry with a
clean, dry cloth.
DO NOT
spray or pour any type of cleaning agent directly onto
the system, as it could lead to injury, such as electrical shock, or
result in damage to the equipment.
To avoid electrical hazards, do not allow fluid of any kind to leak
into the laser console.
Other System Features
HANDPIECE VEIN MODE
The POLYLASE LP software allows both Nd:YAG and Alexandrite handpieces to
operate in
Vein Mode
, i.e. with optional insert objective. The optional VEIN
OBJECTIVE allows to operate handpiece with 3 small spot sizes (2, 3 and 4 mm), which
allow to generate high fluences during spider and varicose veins treatment (specifically,
with Nd:YAG laser). When this optional objective is installed, the user has to switch the
laser into Vein Mode. The handpiece mode selection button is located next to Beam Size
selection window (see Fig. 5-4). The laser is always in
Hair Removal
mode by default
after system start-up, therefore normally this button has name
Hair
displayed on it. In
order to switch to Vein Mode, press and hold
Hair
button for at least 3 sec. The button
will change its name to
Vein
and laser software will switch to
Vein Mode
. The spot size
selection options will change to 2, 3 and 4 mm and all displayed Fluence values will be
re-calculated reflecting small spot size selected. In order to switch the laser back to
Hair
Removal
mode, press and hold
Vein
button for at least 3 sec.
WARNING!
Polylase LP does not have automatic recognition of the optional VEIN
OBJECTIVE. Please, remember to switch the laser into Vein Mode every time you
use the OBJECTIVE. Failure to do so will lead to incorrect Fluences displayed and
may result in patients’ skin burns and/or other injuries.
Содержание POLYLASE LP
Страница 6: ... 2 Figure 1 1 POLYLASE LP Dual Output Laser ...
Страница 26: ... 8 Figure 3 5 Front and side of the POLYLASE LP Laser ...
Страница 32: ... 14 Figure 3 9 Labels rear of the POLYLASE LP Laser Figure 3 10 Handpiece Label Laser Aperture Label 4000 001A ...
Страница 49: ... 10 Figure 5 6 Handpiece protective window is located inside the snap in holder between distance gauges ...