Chapter 3 - How to...
3.4.8.3.5 Disinfection
CAUTION
Wear powder-free gloves at all times when cleaning the device including the scan head and
main mirror. If the gloves become contaminated they must be replaced immediately.
Always follow the cleaning and disinfecting agents’ manufacturer’s instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal to
help assure effective and safe use of the product.
Be careful
NOT
to damage the scan head covers when cleaning.
Do
NOT
use lint cloths, tissues, or other materials that may create dust, on or near the scan
head. Always allow cleaned areas to air-dry.
External surfaces must be cleaned of all visible contamination when they become dirty or soiled with bodily
fluids. Follow your internal clinic or hospital decontamination procedures.
Cleaning of the external surfaces of the device should be carried out on a regular basis, see
3.4.8.4 Battery replacement
CAUTION
Do
NOT
remove the scan head covers. Only Optos trained technical personnel are
permitted to service the device.
A 3V Lithium coin battery is embedded in the hardware for the device and does
NOT
require to be changed by
users. If there is a problem, please contact your Optos representative, see
3.4.8.5 Cable management
WARNING
Do
NOT
plug the scan head into a power block or powered table. The scan head must be
plugged directly into a wall socket.
To prevent electric shock:
l
Only use cables supplied by Optos.
l
Cables should
NOT
be extended or altered.
l
Mains supply must be earthed.
l
The device may only be serviced by Optos trained personnel.
l
Care must be taken never to touch exposed parts of the device while in physical
contact with a patient.
If the mains cable is damaged do not use the system and contact your Optos
representative, see
CAUTION
Ordinary equipment without protection against ingress of water.
You must regularly inspect the equipment. If there is any damage please contact your
Optos representative, see
The device is isolated from the mains by the main power cable and IEC connector. In an
emergency, the device should be isolated by unplugging from the mains power supply.
Ensure the mains socket is easily accessible.
Page 43 of 60
Part
Number:
G102748/11_ENG
Copyright
2021
,
Optos
plc.
All
rights
reserved.
English
Содержание P200T
Страница 1: ...optos com User Guide P200T Optos Panoramic Ophthalmoscope G102748 11_ENG...
Страница 4: ...Page 4 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 10: ...Page 10 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 26: ...Page 26 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 58: ...Page 58 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 60: ...G102748 11_ENG English...