Chapter 1 - Warnings and cautions
1.3 Electrical safety
WARNING
Equipment classification
Electrical Shock Protection - Class 1 Equipment
Type B - relates to the allowable maximum leakage current which can flow from the applied
part. The limits for this are defined in IEC/EN 60601.
The applied parts are the face pad and eye piece.
Medical device. Continuous operation.
Do
NOT
open the scan head covers. There are no user serviceable or replaceable parts
inside.
Do
NOT
plug the scan head into a power block or powered table. The scan head must be
plugged directly into a wall socket.
Do
NOT
disconnect or connect the touch screen while the scan head is in use.
To prevent electric shock:
l
Only use cables supplied by Optos.
l
Cables should
NOT
be extended or altered.
l
Mains supply must be earthed.
l
The device may only be serviced by Optos trained personnel.
l
Care must be taken never to touch exposed parts of the device while in physical
contact with a patient.
If the mains cable is damaged do not use the system and contact your Optos
representative, see
The image server, switch, browser-based review clients and any Viewing PCs or monitors
must be positioned outside the patient environment; more than 1.5 meters (5 ft) from the
scan head.
Do
NOT
connect any peripherals to the scan head. Only use peripherals supplied with the
device. Independently sourced peripherals may
NOT
be compatible. Contact your Optos
representative for more information on compatible peripherals, see
CAUTION
Ordinary equipment without protection against ingress of water.
You must regularly inspect the equipment. If there is any damage please contact your
Optos representative, see
The device is isolated from the mains by the main power cable and IEC connector. In an
emergency, the device should be isolated by unplugging from the mains power supply.
Ensure the mains socket is easily accessible.
Page 14 of 60
English
Part
Number:
G102748/11_ENG
Copyright
2021
,
Optos
plc.
All
rights
reserved.
Содержание P200T
Страница 1: ...optos com User Guide P200T Optos Panoramic Ophthalmoscope G102748 11_ENG...
Страница 4: ...Page 4 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 10: ...Page 10 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 26: ...Page 26 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 58: ...Page 58 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Страница 60: ...G102748 11_ENG English...