APPENDIX A, Service Check Form,S/5 Hemodynamic Modules
A-3(4)
Document No. 800 1008-2
NIBP measurement
S/N
OK
N.A.
Fail
OK
N.A.
Fail
26.
Communication and
memories
27.
Membrane keys
28.
Pump and valves
29.
Leak test
≤
5 mmHg/min
30.
Calibration check
Measured B1
Measured B2
Allowed range
0 mmHg
±
9 mmHg
100 mmHg
100 + z.o.
±
2 mmHg
200 mmHg
200 + z.o.
±
3 mmHg
260 mmHg
260 + z.o.
±
4 mmHg
z.o. = zero offset at 0 mmHg pressure
31.
Watchdog timer activation
pressure
5...10 mmHg
32.
Watchdog timer
Adult
120...140 s
Infant
60...70 s
33.
Safety valve functions
B1
B2
Allowed range
‘Max press’ ADULT
290...330 mmHg
‘2 s after stop’ ADULT
290...330 mmHg
‘Max press’ INFANT
154...165 mmHg
‘2 s after stop’ INFANT
154...165 mmHg
OK
N.A.
Fail
OK
N.A.
Fail
34.
Cuff related messages
35.
Adult cuff detection
36.
Test measurement
37.
Infant cuff detection
Notes