Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
50
TROUBLE
CAUSE
TREATMENT
Out of range
Measured pressure is beyond the
internal measurement range of
the module.
The waveform hits the top and the
numeric display not shown. Check
transducer and its level. Zero the
channel.
4.1.7
Impedance respiration
TROUBLE
CAUSE
TREATMENT
No resp trace
Waveform not selected on the
screen
Press the Monitor Setup key and
make adjustments
Module not plugged in correctly
Plug in
Unacceptable resp waveform
Poor electrode or poor electrode
skin contact
Electrodes from different
manufacturers are used. Too
much/little gel is used.
Poor electrode condition
Electrodes are dried out.
Improper site of electrodes
Check that electrodes are not placed
over bones, active muscles, or layers
of fat.
Improper skin preparation
Remove body hair. Clean
attachment site carefully with
alcohol.
Message: ‘SMALL RESP CURVE’
Respiration signal is very small
With 3-lead cable in ESTPR/NESTPR
try another lead connection I, II, III or
try 5-lead cable.
Message: ‘APNEA ALARM’, and
respiration waveform normal
Respiration source is CO
2
Check respiration source and
change it to correct one.