Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
66
DC gain
shows the gain of DC signal adjusted by the module.
IDC
is the value of infrared signal.
RDC
is the dc value of red signal.
AC gain
is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high
ac gain and 0 means low gain.
Pre gain
is a preamplifier gain for infrared and red signals. Pre gain values can be 1 or 0. Value 1
means normal operation. Value 0 means that signal levels are very low and extra gain is taken into
use.
Temp error
shows the status of the temperature test. No errors found shows the status (OFF) and
errors found (ON).
Protect key
shows normally OFF but turns to ON when the button at the bottom of the module is
pressed.
Protect mode
is normally ON.
It turns to OFF when Protect is switched to OFF for the temperature
calibration in Calibration Menu.
Configuration
shows the chosen module configuration: TP, ST, or STP.
Timeouts
is a cumulative number that indicates how many times the module has not responded to
the monitor’s inquiry.
Bad checksums
is a cumulative number that indicates how many times
communication from the module to monitor broke down.
Bad c-s by mod
is a cumulative number that indicates how many communication errors the
module has detected.
The monitor starts counting these items at power up and resets to zero at power off. The nonzero
values do not indicate a failure, but the continuous counting (more than 50 per second) indicates
either serial communication failure, or module not in place. Also other modules can cause
communication errors that cause these numbers rise.
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum at the EPROM is in accordance with the one the software
has calculated.
EEPROM
indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail
or
?
(module not in place or a communication error).