Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
42
−
To set the protection on:
Turn the toggle switch to the right. Select
Protection ON
and push the ComWheel.
Then turn the toggle switch back to the left.
•
Remove the module from the frame and plug it back again. Then perform
(see the preceding page) to verify the new calibration.
3.4.3
Temperature calibration
NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are
needed. A test set of two plugs is available from Datex-Ohmeda, order code 884515.
Calibrate temperature when measured test values deviate more than
±
0.1 °C, and always after
STP board replacement.
1.
Enter ESTPR: STP service menu.
2.
Enter
Calibrations
menu.
3.
Press the protect button at the bottom of the module and choose OFF in protect mode.
Release the button.
4.
Select
Calibrate T1/Calibrate T2
.
5.
Insert calibration plug (25 °C) into T1/T2 connector.
6.
Press the ComWheel.
7.
Insert calibration plug (45 °C) into T1/T2 connector.
8.
Press the ComWheel.
9.
Press in the protect button at the bottom of the module and choose ON in protect mode.
Release the button.
3.4.4
Invasive pressure calibration
Calibrate invasive pressure when the pressure transducer (probe) is replaced with a different type
of transducer, and when STP board is replaced.
1.
Enter ESTPR: the STP service menu.
(
Monitor Setup
,
Install/Service
(password 16-4-34),
Service
(password 26-23-8),
Parameters
).
2.
Enter
Calibrations
menu.
3.
Connect a pressure transducer with a pressure manometer to the P1/P2 connector. Choose
Calibrate P1
or
Calibrate P2
. Leave the transducer to room air pressure.
4.
Press the ComWheel to start zeroing.
5.
Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
6.
Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel. A tolerance of ±1 mmHg is allowed.
7.
The message ‘Calibrated’ will appear on the display.