S/5 Hemodynamic modules
Document No. 800 1008-2
21
Pulse oximetry measurement section
LED driving circuit
Amplifier
G = 275
IR DC level
Red DC level
Probe
G =1/4096-1
I=5-350mA
Level of LED current
measurement and
feedback circuit
Preamplifier
Current - to - Voltage
converter
Digitally controlled
attenuator
IRed LED intensity
adjustment
Red LED intensity
adjustment
Level of LED current indication (to CPU)
IRed AC signal
Red AC signal
Two step
AC amplifier
G =16
or 63
for NESTPR
G =18
or 125
G =16
or 63
Figure 11
Pulse oximetry measurement block diagram
LED control signals
The processor sends pulse width modulated signals, IRED intensity and RED intensity, which are
converted to DC voltage and filtered. By switches either RED or IRED intensity is sent forward to
amplifier in LED driving circuit.
LED driving circuit
Voltage difference which corresponds to LED current, is measured by the differential amplifier
circuit and its output is sent back to the processor in 0 to 5 V level. There are feedback circuits from
LED current measurement and LED intensity control.
Background light is measured by picking up a sample from the signal. The sample is modified to 0
to 5 V level and sent to the processor.
Measured signal preamplification
Preamplifier is current-to-voltage converter with gain selection. The higher gain is used for
measuring of thin tissue.
Digitally controlled amplifier
D/A converter is a digitally controlled amplifier after which there is another constant amplifier.