Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
62
5.2
ECG service menu
Power freq
Set power frequency; 50 Hz/60 Hz.
Filter low
Set filter low frequency; 0.05 Hz/0.5 Hz.
Filter high
Set filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /100 Hz or 150 Hz @ NE12STPR.
Service Data field
Power freq, and Cable type
show the values chosen or detected,
Filter low and high
defines the
selected filter (Monitor/Diagnostic/ST).
Quick zero
@ NESTPR and ESTPR modules is ON when the signal in any of the three internal
amplifier goes beyond scale, and therefore, a capacitor connected to the related channel
discharges overvoltage. At least one of
Quick zero
values is OFF when 3-lead cable is used. All
three values are OFF when 5-lead cable is used.
Quick zero
also takes place when lead is changed
in 3-lead measurement. @ NE12STPR
Quick zero
is on if any of the ECG amplifiers goes beyond
the scale.
Cable
shows ON when an ECG cable is connected.
Electrode
shows ON when each of these electrodes are connected.
Pacer count
is a running number for pacemaker users.
The front panel ECG key function is confirmed by pressing the key and observing OFF turns to ON at
Button
.
NOTE: M-NE12STPR and M-NESTPR Module does not consist ECG key.
Resp Available
indicates that ECG hardware is capable of measuring impedance respiration.
Measurement
shows ON when the respiration measurement is on.
Amp zero
shows
ON when zeroing of the respiration amplifier takes place.
Waveform
VALUE
will be updated in one second interval.