![Datex-Ohmeda S/5 M-ESTPR Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-estpr/s-5-m-estpr_technical-reference-manual-slot_513843066.webp)
Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
60
5.1.6
NIBP pneumatics menu
Start Pump/Stop Pump
A manual control for the pump. The selection changes to
Stop Pump
when the pump turns on.
Open Exh1/Close Exh1
A manual control for the exhaust valve 1. The selection changes to Close Exh1 when the valve is
opened.
Open Exh2/Close Exh2
A manual control for the exhaust valve 2. The selection changes to
Close Exh2
when the valve is
opened.
Set Valve
With
Set Valve
, the opening of the pulse valve is adjusted between 0 and 255 (0 for fully closed
and 255 for fully open). First push the ComWheel, then turn it to adjust the value on screen and
finally push to set the value.
Reset Clock
Reset Clock
will zero the time on the display.
Pneumatics Data field
See NIBP service menu in chapter
for information on general items
Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
to
AD7
as well as
Timeouts
etc.
Pump, Exh1 Valve
, and
Exh2 Valve
show their states.
Pulse Valve
shows how much the valve is opened (0 to 255) during Valve Setting.
Interval 20 mmHg -> 185 mmHg Checking
Select the
Start pump
at different combinations of the valves open/closed and push the
ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 185
mmHg and displays it. When all the valves are closed, the pump should be able to pump the
pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not
stop without selecting the
Stop Pump
by pushing the ComWheel.
Watchdog BEEP
Connect manometer to the front panel and pump pressure into the module. When the AD5 value