SERVICE PROCEDURES
3.1
General service information
Field service of the hemodynamic modules is limited to replacing faulty printed circuit boards or
mechanical parts. Faulty printed circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed description of the fault.
CAUTION
Only trained personnel with appropriate equipment should perform the tests and repairs outlined in
this section. Unauthorized service may void warranty of the unit.
3.2
Service check
These instructions include complete procedures for service check. The service check is
recommended to be performed after any service repair, however, the service check procedures can
be used also for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (
) which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the performed procedure should be signed in the
check form.
The procedures are designed for monitors with S/5 monitor software of revision 01. However, most
of the procedures also apply to monitors, which contain some other monitor software
type/revision.
3.2.1
Recommended tools
Tool
Order No.
Notes
Patient simulator
-
Pressure manometer
-
Temperature test set
884515
3-lead ECG trunk cable
5-lead ECG cable
10-leadwire ECG cable
SpO
2
finger probe
SAS-F4
InvBP transducer
Adult NIBP cuff & hose
Infant NIBP cuff & hose
Screwdriver