S/5 Hemodynamic modules
Document No. 800 1008-2
39
3.3
Disassembly and reassembly
3.3.1
M-ESTPR, M-ESTR, and M-ETPR modules
Disassemble the M-ESTPR/-ESTR/-ETPR module in the following way. See the exploded view of
1.
Remove the two screws from the back of the module.
2.
Pull the module box slowly rearwards and detach it from the main body. Be careful with
loose latch and spring pin for locking.
3.
To detach the ECG board, remove four screws, disconnect the ribbon cable from the STP
board, and the ribbon cable from the front panel. Slide the board rearward to disconnect
the fixed 10-pin connector from the ECG input board.
4.
To detach the STP board, remove two screws and disconnect the two connectors from the
SP input board. The T-input connector cables must be disconnected as well.
CAUTION
When reassembling the module, make sure that all cables are reconnected
properly.
3.3.2
M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules
Disassemble the M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR module in the
following way. See the exploded view of the module in
1.
Remove the two screws from the back of the module.
2.
Pull the module box slowly rearwards and detach it from the main body. Be careful with
loose latch and spring pin for locking.
3.
To detach the ECG board, detach four screws, disconnect ribbon cable from the STP board
(supply voltage), and ribbon cable from the ECG input board.
4.
When the ECG board is removed, the STP board can be detached by removing four screws,
disconnecting the cable from the membrane keypad, the cable from the temperature
connectors, and cables from the SP input board. Also disconnect the NIBP hoses and the
ribbon cable from the NIBP board.
5.
When the ECG board and the STP board are removed, the NIBP board can be detached by
removing four screws. The joining chamber can be detached by removing three screws and
disconnecting the hoses from the pressure transducers and the pump. The pump can be
detached by removing two screws. If the filter for the air inlet of the pump is removed, it
must be replaced.