‘Probe’ show ON and the corresponding pressure waveform appears onto the screen.
Perform the same check also for the InvBP channel P2.
?
22.
Calibrate the InvBP channels P1 and P2 according to the instructions in the chapter
?
23.
Check the InvBP channels with a patient simulator.
The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
SENSITIVITY -switch - 5 µV/V/mmHg
ECG - BASE - BPM - 60 - BP - 1 - WAVE - ATM
2 - WAVE - ATM
Restore the normal monitoring screen by pressing the key
Normal Screen
.
Connect cables from the channels BP1 and BP2 to the module connectors P1 and P2. Zero
the InvBP channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel.
BP - 1 - WAVE - ART
2 - WAVE - CVP
Check that appropriate InvBP waveforms are shown and the InvBP values are approximately
120/80 (
±
3 mmHg) for the channel P1 and 15/10 (
±
2 mmHg) for the channel P2.
Check that HR- value is calculated from P1 when ECG is not measured (ECG cable
disconnected).
?
SpO
2
measurement
24.
Check that the message ‘No probe’ is shown when no SpO
2
sensor is connected to the
module. Connect a SpO
2
finger probe to the module. Check that the message ‘Probe off’ is
shown when the probe is not connected to a finger.
?
25.
Connect the SpO
2
probe onto your finger. Check that the reading of
95-99 and SpO
2
waveform appear. Check that HR- value is calculated from SpO
2
when ECG
and InvBP (P1) are not measured.
?