S/5 Hemodynamic modules

Document No. 800 1008-2

27

ECG and STP board

There are test pin blocks identical both on STP and ECG boards. Pins and voltages are as follows:

ESTPR

X11

pin 1

+5 Vref

pin 2

+5 V

pin 3

+12 V

pin 4

Gnd

pin 5

-12 V

X12

pin 1

-5 V (STP board only)

X11

X12

NESTPR

X11

pin 1

+5 Vref

pin 2

+5 V

pin 3

+7 V

pin 4

Gnd

pin 5

-7 V

X12

pin 1

-5 V (STP board only)

X11

X12

Содержание S/5 M-ESTPR

Страница 1: ...5TM ESTR Module M ESTR rev 04 S 5TM ETPR Module M ETPR rev 04 Technical Reference Manual Slot All specifications are subject to change without notice Document No 800 1008 2 June 2001 Datex Ohmeda Inc 3030 Ohmeda Drive 53707 7550 MADISON WIS USA Tel 1 608 221 1551 Fax 1 608 222 9147 www us datex ohmeda com Datex Ohmeda Division Instrumentarium Corp P O Box 900 FIN 00031 DATEX OHMEDA FINLAND Tel 358...

Страница 2: ......

Страница 3: ...6 1 3 6 Respiration 6 2 Functional Description 7 2 1 Measurement principle 7 2 1 1 NIBP 7 2 1 2 ECG 7 2 1 3 Pulse oximetry 7 2 1 4 Temperature 9 2 1 5 Invasive blood pressure 10 2 1 6 Respiration 10 2 2 Main components 10 M ESTPR ETPR ESTR modules 10 M NE12STPR NE12STR NE12TPR NESTPR NESTR NETPR modules 11 2 2 3 NIBP board 12 2 2 4 ECG board in 3 and 5 lead measurement 14 2 2 5 ECG board in 12 lea...

Страница 4: ...odule troubleshooting 52 5 Service Menu 53 5 1 NIBP service menu 54 5 1 1 NIBP demo menu 55 5 1 2 NIBP calibration menu 56 5 1 3 NIBP safety valve menu 57 5 1 4 NIBP pulse valve menu 58 5 1 5 NIBP buttons leds menu 59 5 1 6 NIBP pneumatics menu 60 5 1 7 NIBP watchdog menu 61 5 2 ECG service menu 62 5 2 1 ECG setup menu 64 5 3 STP service menu 65 5 3 1 STP calibration menu 67 6 Spare Parts 68 6 1 S...

Страница 5: ...re 4 Front panel of M NESTPR 11 Figure 5 NIBP board functional block diagram 12 Figure 6 3 and 5 lead ECG board block diagram 14 Figure 7 12 lead ECG measurement block diagram 16 Figure 8 STP board block diagram 19 Figure 9 Temperature measurement principle 20 Figure 10 Pressure measurement principle 20 Figure 11 Pulse oximetry measurement block diagram 21 Figure 12 Serial communication and opto i...

Страница 6: ...Datex Ohmeda S 5 monitors Document No 800 1008 2 iv This page intentionally left blank ...

Страница 7: ...NETPR are double width modules designed for use with S 5 monitors The modules provide general hemodynamic parameters Later in this manual modules can be called w o system name S 5 NOTE Do not use identical modules in the same monitor simultaneously The following modules are considered identical M ESTP EST ETP M ESTPR ESTR ETPR M NESTPR NESTR NETPR M NE12STPR NE12STR NE12TPR M MRI MRIP NOTE 12 lead...

Страница 8: ... 1 kg Power consumption about 9 W 1 2 Typical performance 1 2 1 NIBP Oscillometric measurement principle Measurement range adult 25 260 mmHg child 25 195 mmHg infant 15 145 mmHg Pulse rate range accepted 30 250 bpm Measurement interval from continuous to 1h 2h 4h Typical measuring time adult 23 s infant 20 s Initial inflation pressure adult 185 10 mmHg child 150 10 mmHg infant 120 10 mmHg Venous s...

Страница 9: ...ain software ST level range 9 9 mm 0 9 0 9 mV Resolution 0 1 mm 0 01 mV Averaging calculated from 8 QRS SYNCHRONIZATION Direct ECG analog output of ECG 1 V 1 mV Pacer 5 V and 0 5 2 5 ms pulse 30 ms after pacer peak Defibrillator 5 V and 10 ms pulse 35 ms after R point synchronization 1 2 3 Pulse oximetry Measurement range 40 100 Accuracy 100 80 2 digits SpO2 1 SD 1 80 50 3 digits 50 40 unspecified...

Страница 10: ...Invasive blood pressure Measurement range 40 320 mmHg Measurement accuracy 2 mmHg or 5 Zero adjustment range 150 mmHg Calibration range 20 Scales upper limit is adjustable between 10 and 300 mmHg in steps of 10 Lower limit is 10 of selected upper limit below zero Sweep speed 12 5 25 50 mm s DIGITAL DISPLAY Range 40 320 mmHg Resolution 1 mmHg WAVEFORM DISPLAY Range 30 300 mmHg HEART RATE FROM ARTER...

Страница 11: ...nute 5 seconds in infant mode Zeroing to ambient pressure is done automatically Inflation pressure is adjusted according to the previous systolic pressure typically 40 mmHg above If the systolic pressure is not found inflation pressure is increased typically 50 mmHg Max measurement time adult 2 min child 2 min infant 1 min Pressure transducer accuracy is better than 3 mmHg or 2 whichever is greate...

Страница 12: ...ave respiration artifact rejection Accuracy 5 or 2 mmHg whichever is greater Transducer and input sensitivity 5 µV V mmHg 5 VDC 20 mA max current Filter 0 4 22 Hz adjustable Zero set accuracy 1 mmHg Calibration resolution 1 mmHg Zero time less than 15 s Protection against electrical shock Type CF defibrillation proof NOTE The accuracy of the measurement may be different from the specified dependin...

Страница 13: ...al potential produced with electrodes placed on the surface of the body ECG reflects electrical activity of the heart normal abnormal function of the heart effects of anesthesia on heart function effects of surgery on heart function See the User s Reference Manual for electrodes positions and other information 2 1 3 Pulse oximetry A pulse oximeter measures the light absorption of blood at two wave...

Страница 14: ...od is that the accuracy of SpO2 measurement relative to SaO2func can be maintained even at rather high concentrations of carboxyhemoglobin in blood Independent of the calibration method pulse oximeter is not able to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or methemoglobin levels Plethysmographic pulse wave The plethysmographic waveform is derived from t...

Страница 15: ...dded volume of arterial blood Arterial blood Venous blood Tissue Time Incident light 2 GND I I Figure 2 Absorption of infrared light in the finger probe parts layout and schematic diagram 2 1 4 Temperature The temperature is measured by a probe whose resistance varies when the temperature changes called NTC Negative Temperature Coefficient resistor The resistance can be measured by two complementa...

Страница 16: ...e heart and is electrically zeroed The transducer is a piezo resistive device that converts the pressure signal to a voltage The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed 2 1 6 Respiration Impedance respiration is measured across the thorax between ECG electrodes The signal of the respiration is made by supplying current between electrodes ...

Страница 17: ... have separate T and SP input boards 2 2 2 M NE12STPR NE12STR NE12TPR NESTPR NESTR NETPR modules Figure 4 Front panel of M NESTPR The M NESTPR M NETPR and M NESTR modules contain three main PC boards the STP board the ECG board and the NIBP board They work independently Each of these has their own processor and software EPROM The M NE12STPR M NE12TPR and M NE12STR contain three main PC boards The ...

Страница 18: ...f M NE12STR is similar except the ECG board and ECG input board are different 2 2 3 NIBP board ADDRESS DECODER POWER UP RESET CPU 80C51FA RS485 INTERF ADDRESS LATCH RAM INTERNAL WATCHDOG EEPROM Calib ration Data AD CO N VER TER Write protect switch EPROM FRONT PANEL KEYS SOFT WARE CONTROL DATABUS ADDRESS BUS WATCHDOG TIMER X1 JOINING CHAMBER CUFF M EXHAUST VALVE 1 BLEED VALVE CHECK VALVE PUMP EXHA...

Страница 19: ... reset See the section in ECG board s description RS485 communication Watchdog timer The NIBP board is equipped with software independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than preset time As soon as the cuff pressure rises over a specifiedpressure limit timer starts counting The timer is adjusted to stop the pump and ...

Страница 20: ...WER SOURCE DIRECT ECG OPTO COUPLER POWER SOURCE PATIENT OVERLOAD PROTECTION ANALOG SWITCHES to STP Board ISOLATION uP uP uP SUPPLY VOLTAGES HP Filter uP HP Filter uP HP Filter uP SLEW RATE DETECTOR ANTI ALIAS uP HP Filter uP R L F C N LEAD SELECTION LOGIC Controlled by ECG lead selection signals from uP Resp amplifier Sync Rectifier Amplifier DEFIBRILLA TION DETECTOR uP 31 kHz from uP Figure 6 3 a...

Страница 21: ...ier consists of the analog switches which are used for detecting the respiration signal from 31 kHz amplitude modulated raw signal The amplifier stage consists of the differential amplifier and the last amplifier The differential amplifier consists of the operational amplifiers and the components around them This stage is AC coupled on both sides for minimising the offset voltages The last amplifi...

Страница 22: ...oard for voltages 5 VREF 5 V 12 V GND and 12 V 2 2 5 ECG board in 12 lead measurement The 12 lead ECG measurement consists of the functions which are shown in the figure 7 All functions are located in the ECG board except the front panel connector and the ECG input board Front panel connector and ECG input board Input protection and filtering ECG preamplifiers Respiration impedance amplifiers Resp...

Страница 23: ...erential amplifiers ECG amplifiers and baseline restoration The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore the baseline of the signal in the middle of the display after the change of the signal level e g after the change of the DC offset voltage Pacer detection Pacer detection has been made by using two slew rate detector circuits The pacer dete...

Страница 24: ...0 Hz with 60 Hz reject filter The purpose of filtering is to reduce high frequency noise and low frequency e g respiratory movement artifacts Monitor filter is used in normal monitoring Diagnostic filter is used if more accurate diagnostic information is needed ST filter gives more accurate information of ST segment but reduces high frequency noise The high pass filters 0 5 Hz and 0 05 Hz are done...

Страница 25: ...on section Power non isolation section AD converter 8 chn 12 bit IRed measuring RS485 Driver for module reset RS485 Driver for data Power reset Pox LEDs driver Front panel keys Patient isolation transformer Isolation µprocessor unit Figure 8 STP board block diagram Microprocessor unit As processor Intel s 80C196KC 16 is used There are external memories an 8 bit data bus a 16 MHz oscillator an open...

Страница 26: ...ature measurement principle Invasive blood pressure measurement unit Isolated 5 V voltage is supplied to the pressure transducer From the bridge connection a differential voltage which depends on pressure and supplied voltage is calculated see the formula below Uout Uin pressure 5 V where Uin is 5 V Þ Uout 25 V pressure mmHg Pressure amplification is realized in the instrumentation amplifier Gain ...

Страница 27: ...RED intensity which are converted to DC voltage and filtered By switches either RED or IRED intensity is sent forward to amplifier in LED driving circuit LED driving circuit Voltage difference which corresponds to LED current is measured by the differential amplifier circuit and its output is sent back to the processor in 0 to 5 V level There are feedback circuits from LED current measurement and ...

Страница 28: ... additionally there is an auxiliary logic power reset which keeps the reset active for about 500 ms despite the state of reset in the module bus Time constant determines the power up reset time There are components to prevent the module from sending data during reset Data transmission rate is 500 kbps RS485 Driver RS485 Driver Reset Send data Receive data send receive Receive data Senddata send re...

Страница 29: ...Signal is processed on logical high low level even though the outputs of the opto isolators are analog signals in the isolated section Reset line is an open collector type with a pull up resistor Thus the processor is able to use its internal watch dog function Power supply section Isolated supply voltage of the module is developed from 15 Vdirty voltage from the Central Unit Power supply is a swi...

Страница 30: ...n No I O Signal Note 1 I RESET_RS485 2 I 15 VDC 3 I 15 VDIRTY 4 I 15 VDC 5 I O DATA_RS485 6 I O DATA_RS485 7 Ground Shield 8 I RESET_RS485 9 I CTSB 10 O RTSB 11 I RXDB 12 O TXDB 13 Ground Shield 14 I 32 VDIRTY 15 I GroundDIRTY 16 I CTSC 17 O RTSC 18 I RXDC 19 O TXDC 20 ON STANDBY 21 O PWM_ECG 22 RXDD_RS232 23 TXDD_RS232 24 I 5 VDC 25 I 5 VDC Not used Used only by M ESTPR M ETPR M ESTR and M NIBP m...

Страница 31: ...ure connectors P1 P2 6 8 9 0 1 2 4 5 7 3 6 8 9 0 1 2 4 5 7 3 6 8 9 0 1 2 4 5 7 3 Pin No Signal Pin No Signal Pin No Signal 1 Right arm electrode R 1 Feedback resistor 1 Pressure 2 Left arm electrode L 2 Ground 2 Pressure 3 Right leg electrode RL 3 Not Connected 3 Polarisation ground 4 Left leg electrode F 4 Cable shield probe identification ground 4 Polarisation 5 Chest electrode C 5 Probe identif...

Страница 32: ...V floating 1 GND 6 5 Vanalog 10 GND floating 1 2 5 Vdig 12 lead ECG board component side NIBP board There are test pad blocks on solder side X8 and X6 pads and voltages are X7 X8 X5 X6 X8 1 2 9 10 X6 1 2 7 8 X8 X6 Pin No Signal Pin No Signal 1 GND 1 GND 2 WD out 2 A1 output 3 reset 3 5 V 4 5 V 4 5 V ref 5 15 V dirty 5 B1 out A1 input 6 15 V 6 B1 out 7 15 V 7 B2 out 8 8 B2 out 9 10 GND ...

Страница 33: ... pin blocks identical both on STP and ECG boards Pins and voltages are as follows ESTPR X11 pin 1 5 Vref pin 2 5 V pin 3 12 V pin 4 Gnd pin 5 12 V X12 pin 1 5 V STP board only X11 X12 NESTPR X11 pin 1 5 Vref pin 2 5 V pin 3 7 V pin 4 Gnd pin 5 7 V X12 pin 1 5 V STP board only X11 X12 ...

Страница 34: ...rvice check The service check is recommended to be performed after any service repair however the service check procedures can be used also for determining possible failures The procedures should be performed in ascending order The instructions include a check form Appendix A which should be filled in when performing the procedures The mark in the instructions means that the performed procedure sh...

Страница 35: ...e 2 Check external parts the front cover and the front panel sticker are intact all connectors are intact and are attached properly the module box the latch and the spring pin are intact 3 Replace the NIBP pump filter in NE12STPR NE12TPR NE12STR NESTPR NETPR NESTR modules if necessary Reattach the module box and check that the latch is moving properly Switch the monitor on and wait until the monit...

Страница 36: ...cale Auto ECG measurement 6 Enter the service menu Monitor Setup Install Service password 16 4 34 Service password 26 23 8 Parameters Take down the information regarding module software by selecting Scroll Vers and turning the ComWheel 7 Enter the ESTP ECG service menu Check that the Timeouts Bad checksums and Bad c s by mod values are not increasing faster than by 50 per second Check also that th...

Страница 37: ...des show OFF in the service menu and the message Leads Off is shown on the screen Connect all the leads together for example to a suitable screwdriver Check that all the electrodes show ON and the message Asystole appears Disconnect one of the leads and check that the corresponding electrode in the service menu shows OFF within 10 seconds from the disconnection then reconnect the lead Check the re...

Страница 38: ...nt simulator The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator BASELINE IMPEDANCE switch 500 LEAD SELECT switch II RL LL RESP WAVE NORM RATE 20 OHMS 1 0 RATIO 1 1 APNEA OFF SHIFT OFF Check that the RESP waveform is shown and the RR value is 20 5 Change the position of the BASELINE IMPEDANCE switch and check that appropriate RESP waveform and RR value are shown again within ...

Страница 39: ...according to the instructions in the chapter 3 4 3 Temperature calibration 18 Activate the temperature test by selecting Temp Test from the menu and pressing the ComWheel twice When the message Performing temp test disappears from the digit field check that no error messages appear and Temp error shows OFF for both channels in the service menu 19 Check that the module configuration has been set co...

Страница 40: ...P2 to the module connectors P1 and P2 Zero the InvBP channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel BP 1 WAVE ART 2 WAVE CVP Check that appropriate InvBP waveforms are shown and the InvBP values are approximately 120 80 3 mmHg for the channel P1 and 15 10 2 mmHg for the channel P2 Check that HR value is calculated from P1 when ECG is not measured ECG cable disconnected...

Страница 41: ...the NIBP menu Connect a pressure manometer to the NIBP module cuff connector Select Start Pump and press the ComWheel Check that the pump turns on and the pressure inside the tubing system starts to increase Stop the pump by pressing the ComWheel again when the pressure reaches 280 mmHg Highlight Open Exh1 Press the ComWheel and check that the pressure inside the tubing system starts to drop then ...

Страница 42: ...rs B1 and B2 The values should be within 10 mmHg Connect the pressure manometer to the cuff connector and check the calibration with pressures 100 mmHg 200 mmHg and 260 mmHg The zero offset value must be added to the displayed pressure value in order to determine the real pressure Recalibrate the NIBP measurement according to the instructions in the chapter 3 4 Adjustment and calibration Calibrati...

Страница 43: ...d some round object for example a calibration gas bottle Highlight Start Test Start the adult safety valve test by pressing the ComWheel Wait until the pump stops and the pressure is deflated Check the pressure values Max press and 2 s after stop for both transducers All the values should be within 290 330 mmHg Highlight ADULT Press the ComWheel and check that the text changes now to INFANT Select...

Страница 44: ... Screen 36 Keep the cuff on your arm and perform one NIBP measurement Check that the module gives a reasonable measuring result 37 Connect an infant cuff to cuff connector and wrap it around your fingers Start NIBP measurement and check that the module identifies the cuff i e the text Infant appears into the NIBP digit field Cancel the measurement after the cuff identification All modules 38 Perfo...

Страница 45: ...ESTR NETPR modules Disassemble the M NE12STPR NE12STR NE12TPR NESTPR NESTR NETPR module in the following way See the exploded view of the module in 6 1 12 1 Remove the two screws from the back of the module 2 Pull the module box slowly rearwards and detach it from the main body Be careful with loose latch and spring pin for locking 3 To detach the ECG board detach four screws disconnect ribbon cab...

Страница 46: ...The electronics of NIBP pressure measurement is calibrated at the factory Zeroing pressure is automatically maintained by the processor If the zero point of the pressure transducer drifts more than specified an error message is given and the NIBP board should be recalibrated or replaced The calibration can be checked and recalibrated in the NIBP service menu The calibration of the primary pressure...

Страница 47: ...e calibration Turn the switch to the right by for example a sharp pencil This enables menu selection Protection The message Calibration switch ON appears B Select Protection OFF in the Calibration menu and press the ComWheel C Return the toggle switch to the left Menu selection Calibration is now enabled and Protection is disabled When the calibration is enabled a message Calibration not protected...

Страница 48: ...t calibration plug 25 C into T1 T2 connector 6 Press the ComWheel 7 Insert calibration plug 45 C into T1 T2 connector 8 Press the ComWheel 9 Press in the protect button at the bottom of the module and choose ON in protect mode Release the button 3 4 4 Invasive pressure calibration Calibrate invasive pressure when the pressure transducer probe is replaced with a different type of transducer and whe...

Страница 49: ...urement due to patient movements or shivering Weak pulsation message Weak or unstable oscillation pulses due to artifacts accurate diastolic pressure difficult to measure marked arrhythmia marked drop in diastolic pressure diastolic pressure difficult to measure improper cuff position or attachment too few pulses detected weak or unusual blood circulation may give systolic value Check patient cond...

Страница 50: ...surements check cuff positioning Air leakage message 1 Hose or cuff leaking Reasons 1 Replace cuff cuff damaged replace cuff cuff connector damaged replace cuff connector if the fault is in hose connector O ring damaged or missing replace O ring hose double connector damaged replace hose 2 Hose and cuff OK Reasons 2 Connect or replace tube leakage inside the module replace the whole tubing tube di...

Страница 51: ...lt in pressure transducer replace the NIBP board fault in A D converter replace the NIBP board faulty calibration check calibration missing voltages recalibrate Calibration switch on message EEPROM protection switch at the bottom of the module is turned to right Enables setting the protection OFF in the Calibration menu Turn switch to left if you are not going to calibrate Calibration not protecte...

Страница 52: ...o the left at the bottom of the module 5 Calibration not protected only with S STD93 Protect calibration by selecting Protection ON in the NIBP calibration menu 6 ADC error ADC circuit failure Change NIBP board 7 Watchdog time too short Change NIBP board 8 Watchdog time too long Change NIBP board 9 Watchdog activated Change NIBP board 10 EEPROM checksum error memory failure Change NIBP board 11 Au...

Страница 53: ...rs of fat Improper skin preparation Remove body hair Clean attachment site carefully with alcohol Improper bandwidth filter Check filter No ECG trace Waveform not selected on screen Press the Monitor Setup key and make adjustments Module not plugged in correctly Plug in Noise message High frequency or 50 60 Hz noise Isolate noise source 4 1 4 Pulse oximetry SpO2 TROUBLE CAUSE TREATMENT Message NO ...

Страница 54: ...ld Message FAULTY PROBE Probe is faulty Change the probe No SpO2 No waveform selected on screen Check selected SpO2 waveforms by pressing Monitor Setup key and selecting Modify waveforms Wrong configuration setting Check the configuration settings from the ESTPR STP Calibrations menu Monitor Setup Install Service Service Parameters 4 1 5 Temperature TROUBLE CAUSE TREATMENT Message TEMPERATURE ERRO...

Страница 55: ...2 number field Possibly due to pulsating pressure waveform Open the transducer to air and zero the channel Offset is 150 mmHg Open the transducer to air and zero the channel Defective transducer Replace it and zero the channel Calibration failed message Unsuccessful calibrating of P1 P2 number field possibly due to pulsating waveform Turn the transducer to sphygmomanometer and try again zeroing ta...

Страница 56: ...table resp waveform Poor electrode or poor electrode skin contact Electrodes from different manufacturers are used Too much little gel is used Poor electrode condition Electrodes are dried out Improper site of electrodes Check that electrodes are not placed over bones active muscles or layers of fat Improper skin preparation Remove body hair Clean attachment site carefully with alcohol Message SMA...

Страница 57: ...me slot Error message on screen See error code explanation in Service Manual and fix it Do they work Check tubes and connectors Find leak and fix it Check keyboard connector If OK change front panel Check and replace NIBP board OK NO YES YES NO NO NO YES NO NO YES NO YES Key check in Service Menu OK Pump check in Service Menu OK Leak test in Service Menu OK YES NO Check pump connector If OK change...

Страница 58: ...in same slot ECG module ID on screen Check front panel key functions in Service menu Do they work Check front panel unit ECG board and STP board to find culprit Replace if necessary Remove ECG board STP module ID on screen Replace STP board Replace ECG board NO YES NO NO NO YES NO YES NO YES Has SpO2 measurement failed Replace STP board Replace STP board YES NO YES Has Temp measurement failed Repl...

Страница 59: ... 1008 2 53 5 SERVICE MENU 1 Press the Monitor Setup key 2 Select Install Service password 16 4 34 3 Select Service password 26 23 8 4 Select Parameters NIBP NOTE Parameter values in Service Data fields are for reference only on this chapter ...

Страница 60: ...and valves It exists ON or not OFF depending on service menu function The supply voltage can be turned on by selecting the previous Menu and then the desired menu again AD0 to AD7 show the values of each eight channels of A D converter Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor s inquiry Bad checksums is a cumulative number that indica...

Страница 61: ...ement The menu shows the realtime pressure signals that are measured from the NIBP cuff The measurement result is shown in the adjoining digit field Wave Recording Wave Recording is for selecting the recording option If ON is selected the pressure signals are recorded in realtime onto the M REC paper Remove menu Remove menu widens the displayed waveform area Previous Menu The menu can be closed by...

Страница 62: ...ion ON OFF Select OFF when calibrating Protection can be set to ON or OFF only when the toggle switch at the bottom of the module is set to the right Calibration Calibration selection is available only when protection is OFF NIBP calibration can be performed in the NIBP Service menu as follows NOTE Both channels B1 and B2 must be calibrated simultaneously 1 If Protection is ON change it to OFF by ...

Страница 63: ...ety valve menu Start Test Start test is for starting and Stop test is for stopping the Safety Valve test Safety Valve Data See NIBP Service menu in chapter 5 1 for information on general items Pressure Zero Protect handle Calibr prot 15 V power AD0 to AD7 as well as Timeouts etc Max press and 2 s after stop show the measured values at Safety Valve test Safety Valve Test Adult Infant Wrap an adult ...

Страница 64: ...he module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and displays it on the screen The test can be manually stopped by selecting Stop test The valve can be adjusted between 0 and 255 0 for fully closed and 255 for fully open First select Set Valve and push the ComWheel See the pulse valve value and adjust it by turning the ComWheel Then push the ComWheel to confi...

Страница 65: ...FF have no effect on the module Buttons Leds Data See NIBP Service menu in chapter 5 1 for information on general items Pressure Zero Protect handle Calibr prot 15 V power AD0 to AD7 as well as Timeouts etc Buttons Checking The front panel keys function is confirmed by pressing the key and observing OFF turns to ON at Auto On Off and Start Cancel ...

Страница 66: ... zero the time on the display Pneumatics Data field See NIBP service menu in chapter 5 1 for information on general items Pressure Zero Protect handle Calibr prot 15 V power AD0 to AD7 as well as Timeouts etc Pump Exh1 Valve and Exh2 Valve show their states Pulse Valve shows how much the valve is opened 0 to 255 during Valve Setting Interval 20 mmHg 185 mmHg Checking Select the Start pump at diffe...

Страница 67: ...pping the test Watchdog Data field See NIBP Service menu in chapter 5 1 for information on general items Pressure Zero Protect handle Calibr prot 15 V power AD0 to AD7 as well as Timeouts etc Watchdog Interval shows the time the 15 Vdirty stays on during the test Adult watchdog time testing Select Test ADULT and push the ComWheel Watchdog interval starts counting up seconds and keeps on counting a...

Страница 68: ...values is OFF when 3 lead cable is used All three values are OFF when 5 lead cable is used Quick zero also takes place when lead is changed in 3 lead measurement NE12STPR Quick zero is on if any of the ECG amplifiers goes beyond the scale Cable shows ON when an ECG cable is connected Electrode shows ON when each of these electrodes are connected Pacer count is a running number for pacemaker users ...

Страница 69: ...ts counting these items at power up and resets to zero at power off The nonzero values do not indicate a failure but the continuous counting more than 50 per second indicates either serial communication failure or module not in place Also other modules can cause communication errors that cause these numbers rise RAM indicates the state of the RAM memory ROM indicates whether the checksum at the EP...

Страница 70: ...requency artefacts caused by electrosurgery unit but catches the slow changes in ST segment The ST filter is more susceptible to baseline wander than the monitor filter Pacemaker Selects how to display the pacing pulse of cardiac pacemaker The selections are Show Hide ON R and Sensit Hide the pacing pulse is filtered away from ECG data Show the pacer pulse is filtered away from ECG data but the pu...

Страница 71: ...is outside the ranges Cable shows ON when a corresponding cable is connected to the front panel and Probe shows ON when a corresponding probe is connected to the cable Under Value the measured numeric values are displayed simultaneously Pressure values are real time values and shown in mmHg Temperature values are shown in degrees Celsius The front panel STP keys functions are confirmed by pressing...

Страница 72: ...ion Menu Configuration shows the chosen module configuration TP ST or STP Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor s inquiry Bad checksums is a cumulative number that indicates how many times communication from the module to monitor broke down Bad c s by mod is a cumulative number that indicates how many communication errors the modu...

Страница 73: ...e calibration requires accurate test plugs of value 25 C and 45 C Calibration 1 Select Calibrate T1 Calibrate T2 2 Insert the test plug 25 C into the T1 T2 connector 3 Press the ComWheel 4 Insert the test plug 45 C into the T1 T2 connector 5 Press the ComWheel Calibrate P1 Calibrate P2 The functions are for calibrating the invasive blood pressure channels P1 and P2 The calibrations require a press...

Страница 74: ...endation for stock 6 1 1 M ESTP rev 01 M ETP rev 00 M EST rev 00 1 2 3 5 4 13 7 6 14 8 16 17 15 10 11 Item Item description Order no Membrane keypad 879374 1 Module box wide 886168 2 Spring pin 879182 3 Latch 879181 4 Cross recess screw M3x8 black 616215 5 Front panel unit M ESTP rev 01 880337 5 Front panel unit M ETP rev 00 880941 5 Front panel unit M EST rev 00 880946 6 STP board M ESTP rev 01 M...

Страница 75: ...oard M ESTP rev 03 883119 Use 888957 2 883806 6 1 4 M ESTP rev 04 M ETP rev 03 M EST rev 03 Item Item description Order no ECG software 883806 D4 STP software 884342 7 ECG board M ESTP rev 04 886748 Use 888957 8 Front panel unit M ESTP rev 04 05 887153 8 Front panel unit M ETP rev 03 04 887154 8 Front panel unit M EST rev 03 04 887155 15 T input connectors M ESTP rev 04 05 887152 16 EMC cover 8840...

Страница 76: ...x12 628700 15 T input connectors M ESTP rev 04 05 887152 16 EMC cover 884099 17 EMC cover 886818 6 1 7 M ESTPR rev 02 M ETPR rev 02 M ESTR rev 02 Item Item description Order no 7 ECG RESP board M ESTPR rev 02 888956 A front panel unit includes all the connectors and input boards 6 1 8 M ESTPR rev 03 M ETPR rev 03 M ESTR rev 03 Item Item description Order no 6 STP board 891442 use 8002575 8 Front p...

Страница 77: ... 20 Fitting plate 879510 7 NIBP board M NESTPR rev 00 887520 23 Membrane keypad 888242 8 EMC cover 892305 24 Front panel unit M NESTPR M NESTR M NETPR 888273 891748 891772 9 NIBP pump M NESTPR 889993 25 Metal frame 888230 10 EMC cover 892307 26 Cross recess screw M3x8 black 616215 11 NIBP pump filter 57142 27 Cross cylinder head screw M3x6 61721 13 Damping chamber M NESTPR 888240 28 Cross cylinder...

Страница 78: ... board 8002574 includes the STP software 8000717 6 1 12 M NE12STPR rev 00 M NE12STR rev 00 M NE12TPR rev 00 Item Item description Order no 4 ECG board 894284 18 ECG input board 896913 24 Front panel unit M NE12STPR 896910 24 Front panel unit M NE12STR 896911 24 Front panel unit M NE12TPR 896912 33 Screw 61719 6 1 13 M NE12STPR rev 01 Item Item description Order no 5 STP Board 8002574 ...

Страница 79: ...38 FR 880158 880429 880140 Adaptation M ESTP rev 02 Order No M ETP rev 01 Order No M EST rev 01 Order No DE 881955 881958 881961 EN 881953 881956 881959 ES 884200 885043 885044 FR 881954 881957 881960 IT 886753 886754 886755 NL 886044 886046 886045 SV 885857 885856 885859 Adaptation M ESTPR rev 01 03 Order No M ETPR rev 01 03 Order No M ESTR rev 01 03 Order No DA 892207 892209 892208 DE 886966 886...

Страница 80: ...7709 897613 897646 EN 896985 896986 896987 ES 897712 897616 897649 FI 897715 897619 897652 FR 897710 897614 897647 IT 897713 897617 897650 JA 897719 897623 897656 NL 897711 897615 897648 NO 897717 897621 897654 PT 897714 897618 897651 SV 897716 897620 897653 6 1 15 Front panel stickers for S 5 modules Front panel stickers that are related to the Compact Module type and adaptation Adaptation codes ...

Страница 81: ...98486 898452 FI 898426 898489 898455 FR 898421 898484 898450 IT 898424 898487 898453 JA 898430 8000380 898459 NL 898422 898485 898451 NO 898428 898491 898457 PT 898425 898488 898454 SV 898427 898490 898456 Adaptation S 5 M NE12STPR rev 01 Order No S 5 M NE12STR rev 01 Order No S 5 M NE12TPR rev 01 Order No DA 898442 898470 898481 DE 898433 898461 898472 EN 898432 898460 898471 ES 898436 898464 898...

Страница 82: ...on 02 Service Manual p n 882580 ETP and EST Modules revision 01 Service Manual p n 882580 ESTP Module revision 03 04 and ETP and EST Modules revision 02 03 Technical Reference Manual Slot 896620 1 Main manual is 896 624 NE12STPR Module M NE12STPR rev 00 NE12STR Module M NE12STR rev 00 NE12TPR Module M NE12TPR rev 00 NESTPR Module M NESTPR rev 00 NESTR Module M NESTR rev 00 NETPR Module M NETPR rev...

Страница 83: ...APPENDIX A Service check form S 5 Hemodynamic modules Document No 800 1008 2 77 APPENDIX A ...

Страница 84: ...Datex Ohmeda S 5 monitors Document No 800 1008 2 78 This page intentionally left blank ...

Страница 85: ...est not applicable Fail Test Failed All modules OK N A Fail OK N A Fail 1 Internal parts 2 External parts 3 NIBP pump filter 4 Installation 5 Recognition Notes ECG measurement S N 6 Module software serial numbers ECG RESP STP NIBP OK N A Fail OK N A Fail 7 Communication and memories 8 Membrane key 9 Power frequency 10 Cable recognition 11 Lead detection 12 Test with patient simulator Notes ...

Страница 86: ...Fail OK N A Fail 15 Communication and memories 16 Temperature probe detection 17 Calibration check 18 Temp test function 19 Configuration STP ST TP Notes InvBP measurement S N OK N A Fail OK N A Fail 20 Membrane keys 21 Cable and transducer detection 22 Calibration 23 Test with patient simulator Notes SpO2 measurement S N OK N A Fail OK N A Fail 24 SpO2 probe detection 25 Test measurement Notes ...

Страница 87: ...mmHg 100 z o 2 mmHg 200 mmHg 200 z o 3 mmHg 260 mmHg 260 z o 4 mmHg z o zero offset at 0 mmHg pressure 31 Watchdog timer activation pressure 5 10 mmHg 32 Watchdog timer Adult 120 140 s Infant 60 70 s 33 Safety valve functions B1 B2 Allowed range Max press ADULT 290 330 mmHg 2 s after stop ADULT 290 330 mmHg Max press INFANT 154 165 mmHg 2 s after stop INFANT 154 165 mmHg OK N A Fail OK N A Fail 34...

Страница 88: ...meda S 5 monitors A 4 4 Document No 8001008 2 All modules OK N A Fail OK N A Fail 38 Electrical safety check 39 Functioning after electrical safety check 40 Final cleaning Notes Notes Used Spare Parts Signature ...

Отзывы: