56
5.1.2
NIBP calibration menu
Active Leak Test
Wrap an adult cuff around a pipe and connect the cuff to the module. Select the active leak test
(ON). The module automatically pumps a pressure of 260 mmHg into the cuff. Wait for several
seconds until the pressure stabilizes. Then check that the pressure reading does not drop more
than 5 mmHg per minute. If it does, leaking point(s) should be detected and fixed. Cancel the test
by selecting Active leak test OFF.
Calibration Check
After the calibration check is selected (ON), manually pump pressure into the module and make
sure that the same pressure values are shown both on the display and on manometer. Pressure of
both pressure channels B1 and B2 are shown. Note that if the display shows +2 mmHg at zero
pressure and if you 200 mmHg into the module, the display should show +202 mmHg.
Protection
Software calibration protection (ON/OFF). Select OFF when calibrating. Protection can be set to ON
or OFF only when the toggle switch at the bottom of the module is set to the right.
Calibration
Calibration selection is available only when protection is OFF.
NIBP calibration can be performed in the NIBP Service menu as follows:
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
1.
If
Protection
is ON change it to OFF by first turning the toggle switch to the right at the
bottom of the module, which enables the
Protection
selection. Then turn the toggle switch
to the left to enable
Calibration
.
NOTE : Do not disconnect the module from the frame when turning the switch. The module must be
in the frame during the whole procedure.
NOTE: When the switch is at the right, the NIBP field shows an error message ‘Calibration switch
on!’.
NOTE: When calibration is enabled, a message ‘Calibration not protected’ appears.
2.
For proper zeroing to take place, remove the hose from the front panel connector. Select
Calibration
and push the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in the
NIBP message field. After this a pressure bar will appear beside the menu.