Troubleshooting A-ELEC
Document No. 8502856
D-45
Discharge/recharge cycling test
Run the ADU on batteries until it turns itself off and then let it recharge. Observe that the discharge
current for each pack is about 1A and that the charge current is 450 to 500 mA. Repeat this cycle if
necessary. This will level out the differences in the batteries. This test is recommended when new
batteries are installed.
NOTE: When replacing batteries in ADU all 4 must be replaced at the same time. Batteries supplied
by Datex-Ohmeda should always be used in ADU, because we are testing and selecting the batteries
to each others. ADU problems due to locally purchased batteries we do not take any responsibility
for.
Unexpected reset
In a reset condition, the display shows the message ´System failure - Unexpected Reset, please call
for service´.
If an Unexpected reset occurs the Electronic Ventilator and the Fresh Gas Control units continue to
deliver gas as set, but you will be unable to adjust the ventilator settings. Fresh gas flows and
anesthetic agent settings can be adjusted, but the settings are not displayed on the screen.
How the reset system of the ADU works
When an ADU is started the following happens: As soon as power is applied the watchdog (located at
the Digipower board) activates the reset signal to the CPU. When the +5V line rises above 4,65V an
internal timer in the watchdog keeps the reset active for 200 ms to ensure that the CPU will be
properly reset after power up. When the main SW starts to execute it sends a start command to the A-
EV1. If main SW gets a good response from the A-EV1 it continues with the start-up of the ADU.
However if it receives a message from the A-EV1 saying that it is already running, this is interpreted by
the main SW as an error. Main SW should start the A-EV1, if it is already running this is an indication
that the CPU has done a warm start. This will result in the error messages on the screen and prevent
further operations of the ADU. During normal program execution main SW must do a watchdog kick at
least once every second to prevent it from doing a reset of the CPU.
Display description if a reset occurs
When a reset condition occurs, the ADU displays the following:
1.
The screen blanks out.
2.
After approximately 10 seconds, the “Datex-Ohmeda” is shown.
3.
After another 20 seconds, the screen will appear as if in a normal start-up.
4.
After another 20 seconds, the message “System Failure - Unexpected Reset, Please call for
service” appears in red.
Service recommended actions if a reset occurs
•
Check service error log
•
Check Voltages on diagnostic pages
•
Run HW watchdog test from service page if test fails replace.
−
A. Replace the B-CPU board and repeat the test.
−
B. Replace the Digipower board and repeat the test.
•
On the B-CPU2 board check that the SW cassette is correctly installed and that the insulation is
present.
Содержание Anesthesia Delivery Unit S/5
Страница 2: ......
Страница 14: ......
Страница 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Страница 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Страница 76: ...Datex Ohmeda ADU 5 18 Document No 8502852 Figure 24 Transformer unit Figure 25 Transformer unit 2 1 4 6 ...
Страница 80: ...Datex Ohmeda ADU 5 22 Document No 8502852 Figure 30 Compact Circuit A CC1 exploded view 5 1 2 3 4 ...
Страница 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Страница 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Страница 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Страница 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Страница 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Страница 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Страница 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Страница 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Страница 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Страница 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Страница 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Страница 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Страница 148: ......
Страница 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Страница 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Страница 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Страница 211: ...Spare parts A AUF Document No 8502853 A 59 Figure 53 Exploded view of the gas inlet block and the gas labels 4 1 2 3 ...
Страница 229: ...Technical description A FGC1 Document No 8502854 B 3 Figure 3 Technical principles of the A FGC1 ...
Страница 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Страница 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Страница 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Страница 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Страница 307: ...Spare parts A FGC1 Document No 8502854 B 81 Figure 60 Fresh Gas Control Unit A FGC1 1 2 4 3 6 5 7 8 10 11 12 ...
Страница 311: ...Spare parts A FGC1 Document No 8502854 B 85 Figure 64 Parts inside the A FGC1 1 2 3 4 5 6 7 8 10 ii iii i iv 9 ...
Страница 319: ...Spare parts A FGC1 Document No 8502854 B 93 Figure 71 Gas connection unit 1 2 3 4 5 6 7 8 9 10 11 ...
Страница 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Страница 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Страница 362: ...Datex Ohmeda ADU C 36 Document No 8502855 The ventilator calibrating tree Figure 27 The ventilator calibrating tree ...
Страница 400: ...Datex Ohmeda ADU C 74 Document No 8502855 Ventilator diagnostic page layout Figure 42 Ventilator diagnostic page ...
Страница 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Страница 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Страница 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Страница 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Страница 477: ...Troubleshooting A ELEC Document No 8502856 D 47 Trouble shooting PCMCIA Card Figure 57 Troubleshooting PCMCIA Card ...
Страница 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Страница 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Страница 486: ...Datex Ohmeda ADU D 56 Document No 8502856 Figure 63 Transformer unit Figure 64 Transformer unit 2 1 4 ...
Страница 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...