Troubleshooting A-FGC1
B-64
Document No. 8502854
Vaporizer
failure.
FGC safe state
after 2 minutes.
O
2
zero error
O
2
flow measurement zero
point drift 500 ADC. Note: Error
applies from Main SW version
6.1 and 7.1
O
2
flow measurement failure.
Replace the O
2
flow measurement zeroing valve.
Check the O
2
flow measurement unit.
x
Vaporizer
failure.
FGC safe state
after 2 minutes.
Sg zero error
Sidegas flow measurement
zero point drift 500 ADC. Note:
Error applies from Main SW
version 6.1 and 7.1
Sidegas flow measurement failure.
Replace the sidegas flow measurement unit Zeroing
valve.
Check the mixer flow measurement unit, sidegas.
x
Vapouriser
failure
FGC Mixer 0
unstab.
Zero value not stable during
zeroing
1) Leaking zeroing valve.
2) Faulty flow transducers.
1) Check the mixer zeroing valves and flow
transducers
x
Agent % not
delivered
Decr.flow.
FGC Insuffic.
agent
The measured ratio is
continuously smaller than the
set ratio and agent
proportional valve drive is at
maximum (>4000).
1) The required concentration is out of
reach at the current total flow and
cassette temperature.
2) Agent proportional valve is fully open
or stuck closed i.e. not cap-able of
increasing the agent flow due to an
agent channel occlusion or valve failure.
1) Decrease total flow or anaesthetic concentration
2) Check the agent proportional valve operation
and vaporizer maximum output.
x
N
2
O supply
pressure
lost
Measured N
2
O supply pressure
is below 2.4 bar
N
2
O supply pressure has decreased.
1) Increase the pressure.
Fresh gas
unit
failure
Safe state
Safe state: The AA and N
2
O
deliveries are interrupted. The
gas selection valve is switched
to Air position.
Major electronics or flow measurement
failure.
Message is always related to other
failure.
1) See related messages.
x
Message in
Normal
Screen
Message in
diagnostic
page
Error condition
Possible cause
Recommended
action
Err
log
Содержание Anesthesia Delivery Unit S/5
Страница 2: ......
Страница 14: ......
Страница 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Страница 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Страница 76: ...Datex Ohmeda ADU 5 18 Document No 8502852 Figure 24 Transformer unit Figure 25 Transformer unit 2 1 4 6 ...
Страница 80: ...Datex Ohmeda ADU 5 22 Document No 8502852 Figure 30 Compact Circuit A CC1 exploded view 5 1 2 3 4 ...
Страница 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Страница 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Страница 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Страница 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Страница 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Страница 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Страница 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Страница 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Страница 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Страница 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Страница 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Страница 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Страница 148: ......
Страница 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Страница 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Страница 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Страница 211: ...Spare parts A AUF Document No 8502853 A 59 Figure 53 Exploded view of the gas inlet block and the gas labels 4 1 2 3 ...
Страница 229: ...Technical description A FGC1 Document No 8502854 B 3 Figure 3 Technical principles of the A FGC1 ...
Страница 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Страница 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Страница 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Страница 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Страница 307: ...Spare parts A FGC1 Document No 8502854 B 81 Figure 60 Fresh Gas Control Unit A FGC1 1 2 4 3 6 5 7 8 10 11 12 ...
Страница 311: ...Spare parts A FGC1 Document No 8502854 B 85 Figure 64 Parts inside the A FGC1 1 2 3 4 5 6 7 8 10 ii iii i iv 9 ...
Страница 319: ...Spare parts A FGC1 Document No 8502854 B 93 Figure 71 Gas connection unit 1 2 3 4 5 6 7 8 9 10 11 ...
Страница 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Страница 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Страница 362: ...Datex Ohmeda ADU C 36 Document No 8502855 The ventilator calibrating tree Figure 27 The ventilator calibrating tree ...
Страница 400: ...Datex Ohmeda ADU C 74 Document No 8502855 Ventilator diagnostic page layout Figure 42 Ventilator diagnostic page ...
Страница 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Страница 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Страница 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Страница 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Страница 477: ...Troubleshooting A ELEC Document No 8502856 D 47 Trouble shooting PCMCIA Card Figure 57 Troubleshooting PCMCIA Card ...
Страница 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Страница 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Страница 486: ...Datex Ohmeda ADU D 56 Document No 8502856 Figure 63 Transformer unit Figure 64 Transformer unit 2 1 4 ...
Страница 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...