Datex-Ohmeda ADU
4-10
Document No. 8502852
NOTE: The Anesthesia Monitor must be equipped with S-ARK95 or later software.
Configuring the software
Configuration requirements
The Anesthesia Delivery Unit is already factory configured in order to comply with national standards
and most common practices. The unit may be operated without changing these screen default
settings.
Datex-Ohmeda has designed the ADU so that the default settings may be changed to meet hospital
standards. Once changed and saved, the defaults will become active whenever the unit is turned on.
The factory defaults may always be reset if needed.
WARNING
If the factory defaults values are reset, the flow bars order on the display will be
set-up to ISO standard.
If the factory defaults values are reset, the flow bars order on the display will be set to ISO standard,
i.e. that the flow bars order starting from the left side are
O
2
- Air - N
2
O
. This means that countries
using ANSI standard must change the flow bar order starting from the left side on the display to
N
2
O -
Air - O
2
. This is done by entering the
Install
menu (distributor password level is required)
-
Fresh Gas
Setup
and make necessary configuration changes. The colors of the bars will not be changed if the
factory defaults are reset.
NOTE: All the default changes should be performed first and then saved, all at the same time, by
using the Install menu. See “
Saving the Configuration”
at the end of this section.
Always use the default configuration worksheet during a user default configuration and save a copy of
the sheet in the device history records.
System configuration
Check the following items and change configuration if required:
•
Ambient pressure
on the
Install - User config
menu.
•
Primary driving gas for the Ventilator (O
2
or air) on the
Install
menu.
If the factory default settings are changed for these functions, they cannot be reset by the factory
defaults function. Resetting must be done manually.
Ventilator measurements
Ppeak; Pplat; Peep; Pmin
Fresh gas settings
Agent name; Agent % in fresh gas; Total flow
O
2
flow; N
2
O flow; Air flow
Events transferred during system check
System check test results
Gas Delivery: Agent; Gas Delivery: N
2
O; Leak Tests: AUTO
AUTO leak [ml/min]; Leak Tests: MAN; MAN Leak [ml/min]
Number of confirmed checklist items; Bypass Check
Events transferred once a minute
Содержание Anesthesia Delivery Unit S/5
Страница 2: ......
Страница 14: ......
Страница 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Страница 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Страница 76: ...Datex Ohmeda ADU 5 18 Document No 8502852 Figure 24 Transformer unit Figure 25 Transformer unit 2 1 4 6 ...
Страница 80: ...Datex Ohmeda ADU 5 22 Document No 8502852 Figure 30 Compact Circuit A CC1 exploded view 5 1 2 3 4 ...
Страница 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Страница 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Страница 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Страница 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Страница 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Страница 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Страница 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Страница 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Страница 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Страница 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Страница 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Страница 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Страница 148: ......
Страница 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Страница 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Страница 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Страница 211: ...Spare parts A AUF Document No 8502853 A 59 Figure 53 Exploded view of the gas inlet block and the gas labels 4 1 2 3 ...
Страница 229: ...Technical description A FGC1 Document No 8502854 B 3 Figure 3 Technical principles of the A FGC1 ...
Страница 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Страница 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Страница 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Страница 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Страница 307: ...Spare parts A FGC1 Document No 8502854 B 81 Figure 60 Fresh Gas Control Unit A FGC1 1 2 4 3 6 5 7 8 10 11 12 ...
Страница 311: ...Spare parts A FGC1 Document No 8502854 B 85 Figure 64 Parts inside the A FGC1 1 2 3 4 5 6 7 8 10 ii iii i iv 9 ...
Страница 319: ...Spare parts A FGC1 Document No 8502854 B 93 Figure 71 Gas connection unit 1 2 3 4 5 6 7 8 9 10 11 ...
Страница 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Страница 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Страница 362: ...Datex Ohmeda ADU C 36 Document No 8502855 The ventilator calibrating tree Figure 27 The ventilator calibrating tree ...
Страница 400: ...Datex Ohmeda ADU C 74 Document No 8502855 Ventilator diagnostic page layout Figure 42 Ventilator diagnostic page ...
Страница 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Страница 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Страница 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Страница 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Страница 477: ...Troubleshooting A ELEC Document No 8502856 D 47 Trouble shooting PCMCIA Card Figure 57 Troubleshooting PCMCIA Card ...
Страница 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Страница 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Страница 486: ...Datex Ohmeda ADU D 56 Document No 8502856 Figure 63 Transformer unit Figure 64 Transformer unit 2 1 4 ...
Страница 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...