H-Class
87
Advanced Entry Calibration (continued)
Step
Action
Displayed Message
Comment
L
Use the buttons to set the
Gain Number determined
in the previous step. Press
ENTER to enable the
setting.
TRAN SENSOR GAIN
yyy (0 - 31)
18
This example uses Gain
Number 18.
M
Complete a table (see
example below) using new
measurements, as
follows:
(A) Raise the Printhead
Assembly. Place the
label over the Media
Sensor then lower and
latch the Printhead
Assembly. Record the
sensor reading as P.
(B) Raise the Printhead
Assembly. Place the
liner, notch, or mark
over the Media Sensor
then lower and latch
the Printhead
Assembly. Record the
sensor reading as G
(or M).
(C) Raise the Printhead
Assembly. Remove all
media from the Media
Sensor then lower and
latch the Printhead
Assembly. Record the
sensor reading as E.
TRAN SENSOR GAIN
yyy (0 - 31)
18
Where “yyy” is a
numerical value
representing the current
sensor reading.
The re-sampled
values may differ
from those
previously noted.
This is normal; do
not readjust the
Media Sensor.
Selected Gain Table
Gain Number
Paper
Gap (or Mark)
Empty
18 173 42 9
Содержание H-4212
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 6: ......
Страница 16: ...6 H Class ...
Страница 40: ...30 H Class ...
Страница 86: ...76 H Class ...
Страница 142: ...132 H Class ...
Страница 146: ...136 H Class ...
Страница 150: ...140 H Class ...
Страница 158: ...148 H Class ...
Страница 162: ...152 H Class ...
Страница 176: ...166 H Class ...
Страница 180: ......