Product Reference Guide
5-1
Chapter 5
Calibration
A number of situations require the scale to be calibrated. They are:
•
at initial installation of the scanner/scale
•
if the scale cannot be re-zeroed
•
if you change the weight measure from pounds to kilograms or vice-
versa
•
if diagnostics indicate a calibration error
•
the weigh module has been replaced
Follow the procedures on the following pages to ensure that the scanner/
scale will meet Weights and Measurement requirements.
Certification of the scanner/scale’s weighing apparatus is subject to Fed-
eral, State and Local Weights and Measures statutes and is restricted to
authorized government agencies and/or duly registered agents thereof.
Anytime a scale is calibrated, it should be properly sealed with a lead and
wire or paper seal before being placed into service.
It is your responsibility to check with the appropriate authorities in your
area to ensure compliance with pertinent regulations before removing
any official seals or placing a newly calibrated scanner/scale into ser-
vice.
LEGAL NOTE
Содержание Magellan 9500?
Страница 1: ...Magellan 9500 9500 9504 9504 with SmartSentry Product Reference Guide...
Страница 28: ...1 16 Magellan 9500 NOTES...
Страница 86: ...3 16 Magellan 9500 Figure 3 5 Removing Replacing the WRG Window Slide WRG Window Out Remove Vertical Bezel...
Страница 100: ...4 14 Magellan 9500 NOTES...
Страница 116: ...5 16 Magellan 9500 NOTES...
Страница 396: ...6 280 Magellan 9500 NOTES...
Страница 416: ...B 10 Magellan 9500 NOTES...
Страница 426: ...D 2 Magellan 9500 NOTES...
Страница 442: ...E 16 Magellan 9500 NOTES...
Страница 458: ...F 16 Magellan 9500 NOTES...