Datacolor Conditioner Operation Manual / DCCOPMN_RP1 / Rev 3.0 / 7th August 07
page 41 of 51
9.2
Humidifying and dehumidifying system of CONDITIONER
TM
without lighting
unit
The CONDITIONER™ without lighting unit can be operated within an optimized operating range
according to the temperature-humidity-diagram (Figure 10) in a stable manner and without excessive
condensing.
% r.H.
°C
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
Figure 10: Temperature-humidity-diagram of CONDITIONER
TM
without lighting unit
(Hatched area: optimized operating range of the humidity system)
•
Area inside the frame:
Control range of temperature and relative humidity
•
Hatched
area
Control range of temperature and relative humidity without condensation in the inner chamber.
If the set points for temperature or humidity are outside the optimum range, condensation can arise in the
door area.
CAUTION
Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % RH for
a long period
¾
Dry the appliance completely before shut-down:
-
Set the humidity to 0 % RH and switch on humidity switch (5).
-
Set the temperature set point to 60°C for approx. 2 hours (Manual mode).
-
Only then, shut down the unit at the main switch (4).