Datacolor Conditioner Operation Manual / DCCOPMN_RP1 / Rev 3.0 / 7th August 07
page 31 of 51
Table 4: Operating parameters that can be set at CONDITIONER
TM
without lighting unit
Symbol Parameter
Standard
Value
Value
Range
Note
SP Temperature
set point
21.0°C 10.0
to
90.0°C
Temperature set point for the inner chamber of the
CONDITIONER™
rH Humidity
set
point
65.0
%-RH
10.0 to
90.0
% RH
Humidity set point for the inner chamber of the
CONDITIONER™
To avoid an alarm message during operation without
humidity (humidity switch (5) = OFF) set the parameter
„rH“ = 0 % RH
Al.5 Temperature
(absolute
limit)
90.0°C -10.0
to
100.0°C
The safety controller triggers the absolute limit alarm
(chapter 8.1.1 on page 35) if this temperature is
exceeded.
For technical reasons, the absolute limit that is set
should be at least 5°C above the temperature set
point.
If a temperature of 90°C is uncritical for the charging
material, it is recommended to leave the setting of the
safety controller at the default temperature of 90°C.
Otherwise the limit temperature has to readjust every
time when the temperature set point is changed.
PA.H
Password
0
0 to 999 3-digit numeric password for locking the operating
parameter set up (chap. 6.1, page 31)
6.1
Locking of the operating parameter set up
To avoid changing of operating parameters by unauthorized persons, the set up mode of the operating
parameters can be locked.
•
Define 3-digit numeric password via parameter “PA.H” (Table 3 on page 30 or Table 4 on page 31).
•
Press
the and
buttons simultaneously for 3 seconds to move from the basic display to the
menu for selecting the Conditioner’s operation modes. In the upper display the value “OP.H” for
“operating mode hand” is displayed (Figure 7).
•
Press button to enter the “operating mode lock”. In the upper display the value “OP.L” for
“operating mode lock” is displayed (Figure 7). The value “OP.L” can only be selected, if a password is
set via parameter “PA.H” (Table 3 on page 30 or Table 4 on page 31).
•
Press button
to confirm the change of the operating mode and to go back to the basic display
(Figure 5). Changing of operating parameters according to chap. 6 in not possible now.
Figure 7: Locking and unlocking of set up mode of the operating parameters