3 Installation Procedure
Clarity Controls
you want to acquire, name them and possibly check
This Device Starts the
Run in Clarity
/
Clarity Starts This Device
radiobutton. The
Clarity Starts
This Device
option should be checked only when the GC will not start the
acquisition (that means it will not be triggered by the GC Start button,
external start contact or autosampler). Press the
OK
button.
l
On the
Sampler
tab verify the auto detected sampler version and
components, set the
Syringe Size
and
Solvent Wash Mode
.
l
On
Valves
tab verify the identified valves, on the
Aux
tab verify the
detected auxiliaries.
l
Press
OK
. The
Agilent 7890
item
④
will appear in the
Setup Control
Modules
list of the
System Configuration
dialog.
l
Drag the GC icon from the
Setup Control Modules
list on the left side to
the desired
Instrument
tab
⑤
on the right side
⑥
, or use the
button
⑦
to do so.
Note:
In case there is no detector used with 7890 it is necessary to add other
detector to the same Instrument (e.g. DEMO detector available on each
station), otherwise it would be not possible to send the method and start
the GC by
Clarity
.
- 7 -