5.5
General Commands (All Modes) Power Sensor
The following table shows the general commands for the Radi
P
ower
®
RF power sensor. Please
note that every command has to be terminated with a carriage return. The default
communication setting of the serial USB port is 115200,8,N,1.
Command
Reply
Description
“*IDN?”
“DARE!!, RPR20XXY,
version”
Returns the ID of the Radi
P
ower
®
sensor.
“*OPT?”
4
“0” or “010”
Returns the options installed. “0”
is returned if no options are
installed.
“ID_NUMBER?”
“x.x.x.x.x.x.x.x”
Returns the unique id number.
For example:
114.80.79.87.20.0.0.225
“VERSION_SW?”
“2.30”
Returns SW version.
“REBOOT SYSTEM”
“OK”
Reboots the system / Restarts
embedded software.
“RESET”
“OK”
Resets the Radi
P
ower
®
to default
values.
“TEMPERATURE?”
“272”
Returns board temperature in 0.1
degrees.
In this example: 27,2°C
“BAUD <n>”
“OK”
Sets the baud rate, with:
<n> = 0 (for 57600 bps)
<n> = 1 (for 115200 bps, default)
<n> = 2 (for 230400 bps)
<n> = 3 (for 460800 bps)
“BAUD?”
“0”, “1” or “2”
Returns the current baud rate.
“MODE <m>”
“OK”
Sets mode, with:
<m> = 0 (for RMS mode)
<m> = 1 (for max hold, peak)
<m> = 2 (for envelope tracing
mode)
Remark: Only for models RPR2006P and
RPR2018P
“MODE?”
“0”, “1” or “2”
Returns current mode.
Remark: Only for models RPR2006P and
RPR2018P
*
This table continues on the next page.