Copyright © 2008 – 2019 DARE!! Products
Page 12
4.
Switch on the Radi
C
entre
®
system. The new plug-in card will automatically be
detected and initialized by the Radi
C
entre
®
.
5.
Connect the plug-in card to the desired device(s).
The Radi
C
entre
®
will automatically detect the power sensor when it is plugged into
one of the 4 USB slots of the USB1004A plug-in card. Please note that for correct
auto detection of power heads, one must wait at least 10 seconds between un-
plugging and/or re-connecting any individual power sensor. Not taking the above-
mentioned measures into account may result in incorrect power sensor detection.
Please re-start the Radi
C
entre
®
system in order to re-detect all connected sensors
in this situation.
6.
Connect the Radi
C
entre
®
system to your PC using any of the available interfaces of
the Radi
C
entre
®
system; USB, RS232, Ethernet or IEEE-488 (optional).
7.
Place the interlock plug into the interlock connector of the Radi
C
entre
®
. Some cards
are equipped with a local interlock system. For these plug-in cards a separate
interlock plug is supplied.
The hardware installation for the plug-in card is now complete.
The user can control the plug-in card either through the touchscreen on the Radi
C
entre
®
system
(only available for the 2 and 7-slot versions, not for the 1-slot version), or by using the control
commands in combination with an external software package such as the Radi
M
ation
®
EMC test
software.