SERIES 1200 LIQUID TURBINE METER
Table 4-3: Preamplifier power requirements
(continued)
Power requirement
type
Description
Outputs
Powered pulse output:
•
Type: Square wave
•
Frequency range: 0 to 5 kHz
•
Amplitude: 0 to 5 V
•
1000 Ohm internal pull-ups, 20 mA, max.
Variable voltage output:
•
Type: Square wave
•
Frequency range: 0 to 5 kHz
•
Amplitude: 0 to Supply voltage
•
1000 Ohm internal pull-ups, 70 mA, max.
4.3.5
Grounding
Ground the turbine meter electronics internally for safe operation. Connect a wire to the
chassis ground lug located inside the LME as the primary ground. A secondary ground is
located outside of the LME. Refer to
Figure 4-4
and
Figure 4-5
. Digital grounds should
never be connected to chassis ground.
Figure 4-4: External ground lug
A. External ground lug
User manual
Prepare the turbine meter for use
P/N 3-9008-513