Spyder3 GigE User’s Manual PRELIMINARY
DALSA 03-032-10158-04
81
A3.1 Saving and Restoring PRNU and FPN Coefficients
Saving the Current PRNU Coefficients
Saving the Current FPN Coefficients
Purpose:
Saves the current FPN coefficients. You can save up to four sets of
pixel coefficients
Syntax:
wfc i
Syntax Elements:
i
FPN coefficients set to save.
1
= Coefficient set one
2
= Coefficient set two
3
= Coefficient set three
4
= Coefficient set four
Example:
wfc 2
Loading a Saved Set of Coefficients
Purpose:
Loads a saved set of pixel coefficients. A factory calibrated set of
coefficients is available.
Syntax:
lpc i
Syntax Elements:
i
FPN coefficients set to save.
0
=
Factory calibrated pixel coefficients.
1
= Coefficient set one
2
= Coefficient set two
3
= Coefficient set three
4
= Coefficient set four
Example:
lpc 0
Resetting the Current Pixel Coefficients
Purpose:
Resets the current pixel coefficients to zero. This command does
not reset saved coefficients.
Syntax:
rpc
Notes:
The digital offset is not reset.
Purpose:
Saves the current PRNU coefficients. You can save up to four sets
of pixel coefficients
Syntax:
wpc i
Syntax Elements:
i
PRNU coefficients set to save.
1
= Coefficient set one
2
= Coefficient set two
3
= Coefficient set three
4
= Coefficient set four
Example:
wpc 2
Содержание spyder SG-10-01k80
Страница 16: ...PRELIMINARY Spyder3 GigE User Manual 03 032 10158 04 DALSA 16...
Страница 30: ...PRELIMINARY Spyder3 GigE User Manual 03 032 10158 04 DALSA 30...
Страница 45: ...Spyder3 GigE User s Manual PRELIMINARY DALSA 03 032 10158 04 45...
Страница 104: ...PRELIMINARY Spyder3 GigE User Manual 03 032 10158 04 DALSA 104...
Страница 106: ...PRELIMINARY Spyder3 GigE User Manual 03 032 10158 04 DALSA 106...
Страница 109: ...Spyder3 GigE User s Manual PRELIMINARY DALSA 03 032 10158 04 109...