Falcon 4M Camera Manual
DALSA 03-032-20044-01
47
Calibrating Individual Pixels
Purpose:
Sets an individual pixel’s FPN coefficient.
Syntax
sfc x y i
Syntax Elements:
x
The pixel column number from
1
to
2352
.
y
The pixel row number from
1
to
1728
.
i
Coefficient value in a range from
0
to
1023
.
Notes:
The
sfc
command is not available when the camera is using
the factory calibrated coefficients (
csn 0-7
). You must select
the user coefficient set (
csn 8-15
) before you can perform
FPN calibration. An error message is returned if you attempt to
perform FPN calibration when using
csn 0-7
.
Example:
sfc 10 50