29
on, the required isothermal color can be selected, and the temperature values of the lower limit of
the isothermal zone and the upper limit of the isothermal zone can be selected through the digital
roller.
Isothermal temperature setting menu
Alarm temperature setting: set the alarm type first. When the alarm type is off, other
options cannot be set. When the alarm type is set to high temperature alarm, it means that
the alarm will be given when the temperature of the temperature measurement object is
higher than the set alarm temperature; when the alarm type is set to low temperature alarm,
it means that the alarm will be given when the temperature of the temperature measurement
object is lower than the set alarm temperature.
After setting to high or low temperature alarm, other options can be set:
Alarm color setting, when the target temperature exceeds the set alarm temperature, the area
exceeding the alarm temperature will be displayed in the set alarm color. The system provides 9
colors to choose from. When the alarm color is set to "Colorless", it means that the original color of
the image will not be changed.
Alarm temperature setting, the required alarm temperature value can be selected through the
touch screen.
Alarm linkage options, you can click to close, capture, video. If close is selected, there will be
no alarm linkage. If snapshot or video recording is selected, the alarm will be automatically stored
according to the selected type.
Time interval setting, the time interval (seconds) of alarm linkage (capture or video) can be
selected through the touch screen.
Video recording time setting, the time (minutes) of each video recording can be selected