Maintenance Interval
Navigation :
Service menu
→
Maintenance menu
→
Maintenance interval
[value]
By default, the maintenance interval of detectors is set at
13 months.
It may be useful to adapt the maintenance interval under some conditions (specific
atmospheres, set or variable temperature and humidity conditions, etc.). It can be edited to
be any interval ranging from
3
to
26 months
.
Once the maintenance-interval modification has been selected, the current interval
configured in the U•V6 unit is displayed and the cursor flashes on the first digit of the value
in months (the tens).
Maintenance interval
13 months
The user can change the interval value using the
Up
and
Down
arrow keys.
The user can use the
Right
and
Left
arrow keys to move from one digit of the interval value
to the other, with the first digit changing the tens and the second digit changing the units.
The user can select the value chosen for the calibration interval using the
ENTER
key.
Confirmation is requested.
Set interval ?
09 months
The user can
confirm the modification
of the calibration interval by pressing
ENTER
. The
user can cancel the modification and return to the info menu by pressing
ESC
.
It is recommended that you maintain your installation on a yearly basis in order to
guarantee optimal monitoring. The user is therefore
strongly discouraged
from changing
the value of this interval.
The maintenance alert does not indicate a system malfunction but is merely a reminder.
However, renewing the detector-calibration validity may have a significant impact on the
gas-detection function and may cause the unit to not work properly if the detector is not
correctly calibrated any more.
In the event that a battery has been installed, a maintenance alert is also generated
after four years. This interval cannot be modified.