Date & time
[dd/mm/yyyy hh:mm]
Configuration info
Submenu providing the configuration information for
the U•V6 unit
History
View History
allows the user to view the events that have been logged by the unit. These are
given in chronological order.
Calibration Age
The
Calibration Age
menu allows the user to view the time that has passed since the
detectors were last calibrated. This allows the unit to generate a maintenance alert once the
maintenance interval has been exceeded (see
Paragraph 5.2
Outputs test
Navigation :
Service menu
→
Tests menu
→
Outputs test
This item on the
Test Menu
allows the user to temporarily force the outputs to activate, in
order to verify that they are working correctly (relays and siren).
Once the output test option has been selected, the user must select the output to be tested
using the
Up
and
Down
arrow keys.
The status of the selected output is displayed each time.
General fault relay
Not activated
Technical-Fault Relay
Alarm relay 1
Not activated
Alarm Relay 1
Alarm relay 2
Not activated
Alarm Relay 2
Testing an output of the U•V6 unit means that it is activated, along with the servos that are
connected to it (a gas valve or siren, etc.).
This test does not allow the user to disable an output that has already been enabled.