4)
Sensor output voltage setting
( The rated output of Load cell)
Set it with reference to Calibration sheet, enclosed with the sensor.
Press Up key about 3 seconds to be sensor output voltage mode.
->
Press ENT key to be setting.
Press Shift key to choose the required number.
Press Up key to change the number.
Press ENT key to store it.
->
->
Stored value, Here if it is poor, it displays
Press Function key to release from sensor output voltage mode.
Press Function key to return to measure mode from setting mode.
5) Setting the total capacity of the Sensor
Set it with reference to the Calibration sheet, enclosed with the sensor.
Press Up key about 3 seconds, and then press Shift key three time to choose the
total capacity of the sensor.
->
Press ENT key to be setting.
Press Shift key to choose the required number.
Press Up key to change the number.
Press ENT key to store it.
->
->
Stored value. Here if it is poor, it displays
Press Program key to release from the total capacity of the sensor setting mode.
Press Program key to return to measure mode from setting mode.
♣
When decimal point is used for capa setting, ignore and set it with reference to 7-3,
decimal point setting mode.
24