8. Calibration
8-1. Real-weight Calibration
When this mode is
ON
, it enters real-weight calibration mode and is necessary to prepare
the actual weight. Use this mode as much as possible. But if there is any compelling reason,
OFF
the mode and set the constant calibration mode that input EXC voltage and sensor
output value.
1)
Real-weight
Calibration
arrangement
Press Function key over 3 seconds and then Up key to be
->
Press ENT key to choose real-weight calibration mode.
or
off that is displayed.
Press Shift key to choose what you want.
Press ENT key to store it.
->
->
Then it displays either
or
Here if it is poor, it displays
Press Function key to release from real-weight calibration setting mode.
Press Function key to return to measure mode from setting mode.
2) ZERO Calibration
Press Up key about 3 seconds to return to zero calibration mode.
->
Leave the sensor free-load and press ENT key.
->
->
Here if it is poor, it displays
Press Function key to return to measure mode from setting mode.
3) SPAN Calibration
Press Up key about 3 seconds and then Shift key to be SPAN Calibration mode.
->
Press ENT key to enter CAPA setting mode
->
Add weight.
Press Shift key to choose the required number.
Press Up key to change the number.
21