SW 5 : Gain Selection 100 times
SW 6 : Gain Selection 200 times
SW 7 : Gain Selection 400 times
This is BCD code combination(see the below table)
that can establish from 100 to 150times.
SW 8 : Gain Selection 800 times.
※ Note
Sw 5
Sw 6
Sw 7
Sw 8
2
0
2
1
2
2
2
3
1
2
4
8
100 times 200 times 400 times
800 times
ⓐ
SH : Case Ground earthing signal
ⓑ
RX : Receiving data terminal when using RS-232 option.
ⓒ
GN : Ground earthing terminal.
ⓓ
TX : Sending data terminal when using RS-232 option.
ⓔ
VP : DC power supply input terminal. Used when entirely separating external input signal electronically.
ⓕ
C2 : 0Volt Common terminal of Common2 external input signal.
ⓖ
X1 : Operate in case of contact to C2 terminal when using external input for HOLD signal.
ⓗ
X2 : Operate in case of contact to C2 terminal when using external input for Reset signal.
ⓘ
X3 : Operate in case of contact to C2 terminal when using external input for Determination signal.
(When use R1, R2, R3,R4)
ⓙ
X4 : Operate in case of contact to C2 terminal when using external input for ZERO signal.
If it uses op-08, it operates only by print order.
ⓚ
C1 ; Common terminal of Relay output R1, R2, R3, R4.
ⓛ
R2 ; Relay output 1 Contact terminal(HI output of contact by use)
ⓜ
R2 : Relay output 2 Contact terminal(LO output of contact by use)
○
n
R3 : Relay output 3 Contact terminal(OK output of contact by use, Option-06: HH output).
○
o
R4 : Relay output 4 Contact terminal(Option-06: spare, LL output of contact by use).
10