
♣
Gain Setting Table (off = 0, on = 1)
X1
X100
X200
X300
X400
X500
X600
X700
X800
X900
X1000
X1100
X1200
X1300
X1400
X1500
SW5
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
SW6
0
0
1
1
0
0
1
1
0
0
1
1
0
0
1
1
SW7
0
0
0
0
1
1
1
1
0
0
0
0
1
1
1
1
SW8
0
0
0
0
0
0
0
0
1
1
1
1
1
1
1
1
5-2. Function & characteristics of back panel by the terminal
◀
Caution
1. Turn it off when wiring (main power supply)
2. Please use thick cable for GN to prevent it from having impulse voltage and trouble
sulting from surge.
re
C
di
3. Make sure the function of terminal belows and connect required contact input and output.
4. Always connect to more than 200
Ω
load during analog output.
Do not connect induced load.
5. AG terminal is optional, and of AC 110V and 220V, 220V is fixed when manufactured.
onnection of internal jumper makes possible for conversion (Do not operate at one's
scretion).
5-3. Order of use
①
Turn it off, and then connect sensor to external input, external output, power supply, etc.
②
Set Dip-SW as a output value of sensor and a resistance value of gage.
·
If gage is 350
Ω
, Off 10V SW4. If gage is 120
Ω
, On 5V SW4.
·
Gain = Vout/(external supply voltage x sensor output)
For example, when gage is 350
Ω
and sensor output is 1.3024mV/V,
10V/(10V x 1.3024mV) = 767.813
Here, if it outputs over 10V, over flow is occurred. In this case, Gain must be set 700.
·
For the use of external analog output, do the analog zero tracking ( + ) and adjust
e output by using ZE, SP volume on the back.
th
③
Link the sensor and do calibration.
11