Here, if it is poor, it displays
Press Function key to release from EXC-Voltage setting mode.
Press Function key to return to measure mode from setting mode.
3) Gain setting
This setting mode is to inform the CPU of the amplification factor of AMP.
Refer to the table below. If you set Gain to be 1, it must be set 1000, with 1.000mV of sensor
output voltage.
∵
1000
×
1.000mV => 1V/V
♣
Gain Setting Table (off = 0, on = 1), EXC : 10V Standard
Load cell
output(mV/V)
PASS
10.00
5.00
3.33
2.50
2.00
1.67
1.42
SW5 (G100)
0
1
0
1
0
1
0
1
SW6 (G200)
0
0
1
1
0
0
1
1
SW7 (G400)
0
0
0
0
1
1
1
1
SW8 (G800)
0
0
0
0
0
0
0
0
Load cell output
(mV/V)
1.25
1.11
1.00
0.91
0.83
0.77
0.71
0.66
SW5 (G100)
0
1
0
1
0
1
0
1
SW6 (G200)
0
0
1
1
0
0
1
1
SW7 (G400)
0
0
0
0
1
1
1
1
SW8 (G800)
1
1
1
1
1
1
1
1
Press Up key about 3 seconds and then Shift key to choose Gain setting.
->
Press ENT key one time to be setting.
Press Shift key to choose the required number.
Press Up key to change the number.
Press ENT key to store it.
->
->
Stored value. Here if it is poor, it displays
Press Function key to release from Gain setting mode.
Press Function key to return to measure mode from setting mode.
23