For calibration with real-weight, set Zero when free-load, and then enter the known load
d span value to set SPAN.
an
po
se
sa
It uses over 50% of total capacity as a SPAN value as possible to get correct value. It is
ssible to do calibration by using less 50% of total capacity.
·
For calibration with constant value, refer to calibration sheet enclosed to the sensor and
t the sensor output value, EXC voltage, Gain, Capa. Gain and EXC voltage must be set as
me as S/W on the back.
④
When it finishes calibration, press CAL S/W on the back and take note of displaying value
at the moment. Because of this memo there is no need to repeat the complex process of
calibration. For example, if it values 2554 when CAL S/W is on, change the calibration
mode to real-weight calibration mode, set ZERO when free-load, and On the CAL S/W and
enter 2554 as a SPAN value, then set up SPAN.