2410-41, 2410-42 and 2410-43 FEP Maintenance & Support Guide
5-1
Revision 1.0, May 2006
CHAPTER 5
PERIODIC MAINTENANCE
5.1
PERIODIC MAINTENANCE
The only element of the FEP unit requiring periodic maintenance is the air filter, located on the left hand
side of the unit. The filter should be replaced or cleaned every 12 months in normal operating environ-
ments. It should be replaced more often in dusty environments.
The replacement of air filter or fan should be done with power off. The fan and air filter assembly is
located on the left hand side of the FEP unit. Replacement filters can be purchased from Cyclone Micro-
systems, Cyclone part number 370-1303.
Table 5-1. Air Filter Replacement
1
Loosen the four screws retaining the Fan & Air Filter assembly.
2
Remove the Fan & Air Filter assembly.
3
The air flow element slides out the narrow opening in the Fan & Air Filter
assembly. (see Figure 4-1)
4
Slide a new filter element into the narrow opening in the Fan & Air Filter
assembly.
5
Install the Fan & Air Filter assembly.
6
Snug the four retaining screws using a Phillips screwdriver.