Shenzhen Zhiyong Electronics Co., Ltd.

 

 

Preface 

First of all, thank you for purchasing our products, the product specification is about the 
function,  usage,  operation  attention  points,  etc.  Before  use,  please  read  the  instructions 
carefully and used correctly. 

Manual annotation will use the following symbols to distinguish. 

 

 

 

 

 

 

 

 

 

 

 
 
 

 

 

To avoid short out and deadly accident, the circuit under test should not surpass 600VAC 

 

This probe is not designed for naked conductor. 

 

Do not touch the conductor under test and the transducer probe during measuring. 

 

When the oscilloscope is connected to other test terminal, please be careful for these 

points: 

 

Please use the fundamental insulation equipment with fitful voltage range and 

pollution rate while the probe’s test terminal is connected to the other terminals. 

 

Please do not surpass the safe voltage range if the fundamental insulation of the test 

terminal cannot be met. 

 

Please operating under correct guidance about electric safety.   

 

There could be deadly electric shock if the device is damp or the users’ hand is wet. 

 

 

 

   

Warning 

Notice 

Warning

=n 

In the case of wrong operation, the user has the threat of injury, it records the 
relevant matters needing attention to avoid such dangers. 

The user may have suffered minor injuries and material damage while with the 
wrong operation, to avoid such situation, note the matters needing attention. 

This symbol means it has a harm to the machine and human body, you must 
refer to manual operation. 

 

Note 

Recording of important notes while using the machine. 

CYBERTEK

Содержание CPL8100A

Страница 1: ...Shenzhen Zhiyong Electronics Co Ltd www cybertek cn AC DC Current Probe CPL8100A 100A 600kHz CPL8100B 100A 2MHz Shenzhen Zhiyong Electronics Co Ltd CYBERTEK...

Страница 2: ...n equipment with fitful voltage range and pollution rate while the probe s test terminal is connected to the other terminals Please do not surpass the safe voltage range if the fundamental insulation...

Страница 3: ...th dust and water The probe s upper and lower contact surfaces is made by precise polishing process Please protect this part or the probe could be malfunction When the battery voltage is lower than 6...

Страница 4: ...CPL8100A B Brief Summary 2 Summary 4 Introduction about probe and accessories 4 Electronic Characteristics 6 Operating Method 8 Mechanic Characteristics 9 Environment Characteristics 9 Maintenance 9...

Страница 5: ...the current value The auto zero function make it easy for the users to operate and it also has power and low voltage indicator as well as overload sound alert The probe can be powered by battery or ex...

Страница 6: ...urement users need to do zero degaussing before measuring to avoid the influence of environment factors like earth magnetic field and temperature shift Auto Zero Indicator when Auto Zero the indicator...

Страница 7: ...Figure 3 Phase shift DC 65Hz 1 5 DC 65Hz 1 Typical DC linearity The typical DC linearity at H level 0 01V A Figure 4 Rise time CPL8100A 583ns CPL8100B 175ns Max operation current Please refer to volt...

Страница 8: ...g voltage V H 0 01V A L 0 1V A DC 100 10 600 600 DC AC peak value 100 10 600 600 AC peak value 100 10 600 600 AC peak to peak value 200 20 1200 RMS CAT III 70 7 7 07 300 300 RMS CAT II 70 7 7 07 600 6...

Страница 9: ...the oscilloscope from voltage to current display If user s oscilloscope doesn t have such function user have to manually set the corresponding attenuation ratio for instance if the probe is H level 0...

Страница 10: ...contrary the result is negative Adjust the vertical sensitivity of the oscilloscope to get a stable waveform 5 Mechanic Characteristics Size of the front current clamp About 100 20 60mm Size of the b...

Страница 11: ...e of the product there can be one check or calibration every year 8 Troubleshoot Problem Reason Treatment No output signal or amplitude is low Power is off Turn on the power Oscilloscope is AC mode Pl...

Страница 12: ...Room A1702 Building 4 TianAn Cyber Park HuangGe North Road LongGang District ShenZhen City China Tel 86 400 852 0005 86 755 86628000 Q Q 400 852 0005 Fax 86 755 86620008 Email cybertek cybertek cn Zh...