Shenzhen Zhiyong Electronics Co., Ltd.
2
The probe is consist with precision components such as Magnetic core and Hall sensor,
which could be possibly damaged because of the sudden temperature change or external
force impact. Please be careful for vibration or impact.
The product is neither waterproof nor dustroof, please do not operate in the environment
filled with dust and water.
The probe’s upper and lower contact surfaces is made by precise polishing process.
Please protect this part or the probe could be malfunction.
When the battery voltage is lower than 6.5V, the probe will indicate a low voltage alert,
please change the battery . You can use a 9V alkaline battery for longer time.
CPL8100A/B Brief Summary
Type
Max current
(DC+Pk
)
Max RMS
value
Band width
(-3dB)
Range
Current
transfer
ratio
CPL8100A
100A
70.7 Arms
600kHz
100A
0.01V/A
10A
0.1V/A
CPL8100B
100A
70.7 Arms
2MHz
100A
0.01V/A
10A
0.1V/A
Attention
Note