Shenzhen Zhiyong Electronics Co., Ltd.

 

 

 

According  to  the  value  of  the  current  under  test,  choose  the  correct  range  by  pressing 

button. 

Attention: Different range    correspond to different attenuation ratio of the oscilloscope.   

 

Press  the  Auto  Zero  button  .  The  buzzer  will  beep  for  twice,  or  it  will  have  a  long  and 

continuous beep meaning the failure of Auto Zero .The external magnetic field can have a 

little influence to the DC zero, so do not move the device after Auto Zero. 

 

Open the clamp of the probe and clamp the conductor under test. 

Attention: There’s direction tip on the current probe clamp. If the direction of the current 

fits the tip, output will be positive; On the contrary, the result is negative. 

 

Adjust the vertical sensitivity of the oscilloscope to get a stable waveform. 

 

5. Mechanic Characteristics 

Size of the front current clamp 

    About 100*20*60mm 

Size of the back output box 

    About 137*33*35mm 

Operation altitude

 

    0~2000 meter

 

Maximum diameter of the conductor 
under test

 

    12mm ø

 

Length of the cable connecting 
current clamp and output box 

    1 meter 

Length of double terminal BNC cable

 

    1 meter

 

Weight

 

    About 223g(without battery)

 

6. Environment characteristics 

Operating temperature

 

    0

℃~

+50

 

Storing temperature 

 

    -20

℃~

+80

 

Operating relative humidity 

 

    0

 to +40

℃,

95%RH

 

    +40

~+50

℃,

45%RH

 

Pollution Degree

 

Level 2

 

7. Maintenance 

If the product is in the warranty period and used correctly, and at the same time 

the  malfunction  is  caused  by  the  flaws  of  product  itself  and  have  not  been 

disassemble, the company will give free repair to the product.   

CYBERTEK

Содержание CPL8100A

Страница 1: ...Shenzhen Zhiyong Electronics Co Ltd www cybertek cn AC DC Current Probe CPL8100A 100A 600kHz CPL8100B 100A 2MHz Shenzhen Zhiyong Electronics Co Ltd CYBERTEK...

Страница 2: ...n equipment with fitful voltage range and pollution rate while the probe s test terminal is connected to the other terminals Please do not surpass the safe voltage range if the fundamental insulation...

Страница 3: ...th dust and water The probe s upper and lower contact surfaces is made by precise polishing process Please protect this part or the probe could be malfunction When the battery voltage is lower than 6...

Страница 4: ...CPL8100A B Brief Summary 2 Summary 4 Introduction about probe and accessories 4 Electronic Characteristics 6 Operating Method 8 Mechanic Characteristics 9 Environment Characteristics 9 Maintenance 9...

Страница 5: ...the current value The auto zero function make it easy for the users to operate and it also has power and low voltage indicator as well as overload sound alert The probe can be powered by battery or ex...

Страница 6: ...urement users need to do zero degaussing before measuring to avoid the influence of environment factors like earth magnetic field and temperature shift Auto Zero Indicator when Auto Zero the indicator...

Страница 7: ...Figure 3 Phase shift DC 65Hz 1 5 DC 65Hz 1 Typical DC linearity The typical DC linearity at H level 0 01V A Figure 4 Rise time CPL8100A 583ns CPL8100B 175ns Max operation current Please refer to volt...

Страница 8: ...g voltage V H 0 01V A L 0 1V A DC 100 10 600 600 DC AC peak value 100 10 600 600 AC peak value 100 10 600 600 AC peak to peak value 200 20 1200 RMS CAT III 70 7 7 07 300 300 RMS CAT II 70 7 7 07 600 6...

Страница 9: ...the oscilloscope from voltage to current display If user s oscilloscope doesn t have such function user have to manually set the corresponding attenuation ratio for instance if the probe is H level 0...

Страница 10: ...contrary the result is negative Adjust the vertical sensitivity of the oscilloscope to get a stable waveform 5 Mechanic Characteristics Size of the front current clamp About 100 20 60mm Size of the b...

Страница 11: ...e of the product there can be one check or calibration every year 8 Troubleshoot Problem Reason Treatment No output signal or amplitude is low Power is off Turn on the power Oscilloscope is AC mode Pl...