PROCEDIMIENTO REDOX (mV)
Put in short circuit the BNC connector. Using a
metal wire, connect the core pin with the external
cylinder part.
•
POINT1: Adjust the value on the display
up to reach 0 using + and – buttons
•
Press OK to save the data
•
On the display will appear POINT 2
•
Dip the electrode in a buffer solution. We
suggest 250, 475 or 650mV
•
Adjust the value on the display up to
reach the buffer solution value using +
and - buttons.
•
Press OK to save the data.
ALARMS setting
Overdosing alarm: For any kind of problems may occur in the plan (wrong calibration, dirty or broken probe, etc.) if the
pump doesn't reach the Setpoint in OVER ALARM TIME, the pump stops the dosing. On the display Alarm message will flash,
Alarm Led will flash.
To enter in ALARM menu follow the procedure
described in paragraph - PUMP SETUP (pag.81).
When ALARM menu is reached, press OK to enter
in submenus
OVERDOSING ALARM setting
On the display will appear OVER and time counter
(h:min) will flash. Press + and - to adjust the time,
over which the pump will go in alarm.
Press OK to confirm and to save data.
This will end the alarm setting.
Press ESC to return in measure mode.
84
ENGLISH
Содержание BOMBAPRO PH-RX
Страница 2: ......
Страница 7: ...Limpieza y mantenimiento del electrodo 30 DIBUJOS 117 Bomba 117 Cuerpo bomba 118 ESPAÑOL 7 ...
Страница 32: ......
Страница 35: ...DISEGNI 117 Pompa 117 Corpo Pompa 118 ITALIANO 35 ...
Страница 60: ......
Страница 63: ...DRAWINGS 117 Pump 117 Pump head 118 ENGLISH 63 ...
Страница 88: ......
Страница 91: ...DESSINS 117 Pompe 117 Corps de pompe 118 FRANÇAIS 91 ...
Страница 116: ......
Страница 119: ......
Страница 120: ...CDMU 00203ML1A 03 2016 ...