TECHNICAL DATA
Measuring range
-
80…5
0 °Ctd pressure dew point resp. dew point in °Ctd
0…100 % RH
-
20…70 °C
Type 0699.0510, FA 510
-
80…20 °Ctd
≙
4…20 mA
Type 0699.0512, FA 510
-
20…50 °Ctd
≙
4…20 mA
Other scales on request, for example, -60 ... 30 ° Ctd
ؙ
4 ... 20 mA
Accuracy:
typical
r
1 °Ctd von 20…
-20 °Ctd
r
2 °Ctd von -
50…
-20 °Ctd
r
3 °Ctd von -
50…
-80 °Ctd
Pressure range:
-
1…50 bar
standard
Power supply:
24V VDC (10..30 VDC)
Output:
4…20 mA 3
-wire technology
**
RS 485 (Modbus RTU)
**
Protection class:
IP 65
EMV:
DIN EN 61326
Operating temperature:
-2
0…70 °C (ideal 0…50 °C)
Storage temperature:
-
40…80 °C
Load for analogue output:
< 500 Ohm
Screw-in thread:
G 1/2" stainless steel
Optional: UNF 5/8” or NPT ½”
Material of housing:
zinc alloy
Sensor protection:
sinter filter 50
P
m stainless steel
Connection:
M12, 5-pole
**
Remark:
Parallel use of analogue
4…20mA