INTRODUCTION 

 

Dear Customer, 
 
thank you for choosing the FA 510. Please read these installation and operating instructions 
carefully before installation and commissioning and follow our instructions. The proper 
functioning of the FA 510 and safe operation can only be ensured if the regulations and 
instructions described are strictly observed. 

 
 

Intended Use 

 

 

The dew point sensor is intended for measuring the dew point or the pressure dew point in 
clean, dry and oil-free gases and compressed air.  
 
The user must check whether the instrument is suitable for the selected application. It must 
be ensured that the medium is compatible with the wetted parts. The technical data listed in 
the data sheet are obligatory. 
 
Improper handling or operation outside the technical specifications is not permitted. Claims of 
any kind based on improper use are excluded. 

Содержание FA 510

Страница 1: ...with 3 wire technology 4 20 mA and RS 485 Modbus output enables a reliable and long term stable monitoring of the dew point in industrial applications such as in compressed air plants refrigerating a...

Страница 2: ...ctions The proper functioning of the FA 510 and safe operation can only be ensured if the regulations and instructions described are strictly observed Intended Use The dew point sensor is intended for...

Страница 3: ...will prevent soiling of the FA 510 x Standing air leads to long measuring times DESCRIPTION The FA 510 dew point sensor enables a reliable and long term stable monitoring of the dew point in industria...

Страница 4: ...von 50 80 Ctd Pressure range 1 50 bar standard Power supply 24V VDC 10 30 VDC Output 4 20 mA 3 wire technology RS 485 Modbus RTU Protection class IP 65 EMV DIN EN 61326 Operating temperature 20 70 C i...

Страница 5: ...in 3 Pin 4 Pin 5 FA 510 Connector plug VB RS485 A VB RS485 B I Current output Connection cable 0554 0104 5 m 0554 0105 10 m brown white blue black grey VB Positive supply voltage 24VDC 10 30 VDC smoot...

Страница 6: ...uick coupling In case of compressed air containing oil and dirt particles a pre filter should be installed in front of the measuring chamber Compressed air flows continuously at 7 bar approx 1 l min e...

Страница 7: ...er x Function code 16 Write multiple Register Register Mapping measuring values Modbus Register Modbus Address No of Byte Data Type Description Defau lt Settin g Read Write Unit Comment 1001 1000 4 Fl...

Страница 8: ...isabled 1 Modbus Enabled Analog Scaling Settings 2007 2011 Modbus Register Modbus Address No of Byte Data Type Description Default Setting Read Write Unit Comment 2007 2006 4 UInt32 Output Value 4 R W...

Страница 9: ...dition to the warranty service necessary repairs adjustments or similar are carried out the warranty services are free of charge but there is a charge for other services such as transport and packing...

Страница 10: ...ove mentioned components comply with requirements of the following EU directives Elektromagnetische Vertr glichkeit Electromagntic compatibility 2014 30 EUG 2014 30 EC RoHS Restriction of certain Haza...

Страница 11: ...tr 15 D 78052 VS Tannheim Tel 49 0 7705 978 99 0 Fax 49 0 7705 978 99 20 Mail info cs instruments com Web http www cs instruments com Gesch ftsstelle Nord Sales Office North Am Oxer 28c D 24955 Harris...

Отзывы: