Component Removal and Replacement
4-26
Service Manual
Note:
On older model ventilators, the oxygen sensor may be screwed in; if so, remove by turning it
counterclockwise.
3.
Reinstall the oxygen sensor by reversing steps 1 and 2.
4.19
Inhalation Outlet Block and Mixing Block Assembly
Figure 4-21.
Inhalation Outlet Block and Mixing Block
To remove and reinstall the inhalation outlet block and the mixing block, refer to
perform the following steps:
1.
Remove the oxygen sensor by performing the procedure in section
.”
2.
Remove the two screws on top of the mixing block (BLK2107M).
3.
Remove both the mixing block and inhalation block (BLK2119M) from the ventilator.
4.
Remove the two screws and washers securing the mixing block to the inhalation outlet block.
5.
Separate the mixing block from the inhalation outlet block.
6.
Reinstall the inhalation outlet block and mixing block by reversing steps 2 through 5.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......