Oxygen Sensor (SEN2103P)
Service Manual
4-25
Figure 4-19.
Outlet Block (BLK2108M)
3.
Pull BLK2108M away from the ventilator, and remove the two tubes.
4.
Reinstall the outlet block by reversing steps 2 and 3.
4.18
Oxygen Sensor (SEN2103P)
Figure 4-20.
Oxygen Sensor (SEN2103P)
To replace the oxygen sensor, refer to
,and perform the following steps:
1.
Remove the oxygen sensor cable (GR-CBL2109A) from the oxygen sensor (SEN2103P) by unscrewing the
connector on the cable.
2.
Remove the oxygen sensor by pulling it out.
1
Outlet block
(GR-BLK2108A)
3
Oxygen sensor cable
(GR-CBL2109A)
2
Screws