Note

Attaching and Removing the Gas Inlet Tube

●Attaching

Insert the tube into the coupling on the Gas Detector inlet and secure it in place 
with the locking tab to seal the edge of the tube. After the tube is attached, make 
sure that it will not pull out.

●Removing

Press in the release ring on the coupling and simultaneously pull out the gas inlet 
tube.

Attaching                                                         Removing

Gas Inlet Tube

●Filter case

Contains the filter (FE-112).

●Gas inlet tube

One meter tube that connects to the 

Gas Detector

●Gas sampling pipe

    This is the gas inlet.

Tube adapter

Release ring

●Tube adapter

connects to the gas inlet on the 

main unit

4

5

No

Name

Function

Power switch

Press and hold to turn the Gas Detector ON and OFF.

A l a r m   s e t t i n g  

indicator switch

Used to check the alarm settings. Press the switch repeatedly to 

display the first level alarm setting, second level alarm setting, and gas 

concentration.

Buzzer stop switch

Zero calibration switch Press and hold to adjust the zero point automatically.

CN jack

Connects to the plug of a CA-7 Communications Adapter (option) used 

for log data collection.

Special command switch Used with the logger function.

UP/DOWN switches Used to set the start time and other logger function settings.

Key Switches

LCD Display

●Log indicator

    Lit when the log function is operating.

●Flow rate drop

    Lit when the flow rate 

drops.

●Gas concentration alarms

Lit when the detected gas concentration reaches the 

alarm point.

●Sensor trouble

    Lit when a sensor error occurs.

●Gas sampling flow rate

    Normal: RUN, Error: STOP

●Gas concentration bar graph

    Each bar is 5% of full scale. Bars lit all the 

way to the right indicate full scale.

●Gas concentration

    Indicates the gas concentration and unit.

●Battery indicator

    Indicates the battery capacity.

Full

 ・
 ・
 ・

Exhausted (flashing)

①  ②  ③  ⑤

                    

Содержание XPS-7

Страница 1: ...his manual Instruction manual No XPS 7T 10 Semiconductor Plant Gas Detector XPS 7 Instruction Manual Store this Instruction manual in a convenient location and consult it whenever necessary Operate th...

Страница 2: ...ing Procedure 6 5 1 Inserting Batteries 6 5 2 Installing the Sensor Unit 7 5 3 Turning the Gas Detector ON 8 5 4 Detection 8 5 5 Ending Detection 8 6 Logger Function 9 7 Replacing the Filter 10 8 Dail...

Страница 3: ...ath WARNING Indicates a potentially hazardous situation that if not avoided could re sult in serious injury or death CAUTION Indicates a potentially hazardous situation that if not avoided could re su...

Страница 4: ...es FS Alarm setting 1st level 2nd level XDS 7NH NH3 Ammonia 100 ppm 12 ppm 25 ppm XDS 7SH SiH4 Silane 25 ppm 2 5 ppm 5 ppm XDS 7DC SiH2Cl2 Dichlorosilane 25 ppm 2 5 ppm 5 ppm XDS 7AH AsH3 Arsine 250 p...

Страница 5: ...to open it and access the key switches Buzzer speaker Gas inlet Sampled gas aspiration inlet 6 dia Attaches to the gas inlet tube Gas outlet Exhaust outlet for sampled gas 6 dia Adapter connector Con...

Страница 6: ...ecial command switch Used with the logger function UP DOWN switches Used to set the start time and other logger function settings Key Switches LCD Display Log indicator Lit when the log function is op...

Страница 7: ...is attached make sure that it will not pull out Removing Press in the release ring on the coupling and simultaneously pull out the gas inlet tube Attaching Removing Gas Inlet Tube Filter case Contains...

Страница 8: ...AA alkaline batteries provided and make sure they are inserted according to the polarity indicated in the battery compartment Close the battery cover WARNING Replace all four batteries at the same tim...

Страница 9: ...power to the sensor unit A sensor unit that does not have power to it cannot detect gases and may cause the sensor trouble indicator to light A sensor unit installed in the Gas Detector is powered eve...

Страница 10: ...after a power failure 5 4 Detection Place the gas sampling pipe near the target detection area When the unit detects gas the concentration of the detected gas is displayed If the detected gas concentr...

Страница 11: ...e log data to a personal computer Press the DOWN switch once in RUN mode LOG will be displayed on the lower right corner of the LCD display and the log start time can be entered Press the SPECIAL COMM...

Страница 12: ...re factory adjustments that are not available for user application 7 Replacing the Filter Twist the lter holder with detection probe to open it Use a pair of tweezers to remove the O ring and filter I...

Страница 13: ...age 6 The batteries are low Replace the batteries The sensor unit is not installed Install a sensor unit Page 7 Abnormal drop in the ow rate The lter is clogged Replace the lter with a new one Page 10...

Страница 14: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation s...

Страница 15: ...2 hours of continuous operation at 20 with alkaline batteries and no alarms Operating temperature range 0 to 40 Dimensions W 62 x H 150 x D 128 mm projected portions excluded Mass Approx 1 3 kg Approv...

Страница 16: ...Distributor Manufacturer New Cosmos Electric Co Ltd 2 5 4 Mitsuya naka Yodogawa ku Osaka 532 0036 Japan URL http www newcosmos global com XPS 7 09 T XPS 7T 10 2008 100...

Отзывы: